Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000819

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-08-20 06:50 83831 639228054237135192-6185 S0003591 Fail View
2026-08-20 06:51 83832 639228054645693156-4169 S0003591 Fail View
2026-08-20 07:08 83847 639228064559227618-7731 S0003591 Pass View

Test 83831

Result
Fail
Date
2026-08-20 06:50
Article
S0003591
ID record
639228054237135192-6185
ID production
Temperature
Humidity
Test duration
14
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5598.23368 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 998.757143 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.455266987 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.691471658 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1252298.41 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.14002625 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.28819729 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28190686 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.28692015 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79721847 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9968382 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00804081 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 12.0310636 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-08-20 06:50 998.757 950 1050
2026-08-20 06:51 998.807 950 1050
2026-08-20 07:08 998.704 950 1050