Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000750

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-08-19 10:10 83344 639227309946882470-7354 S0003591 Fail View
2026-08-19 10:11 83345 639227310338075339-3139 S0003591 Fail View
2026-08-19 10:12 83347 639227310926898359-4482 S0003591 Pass View

Test 83344

Result
Fail
Date
2026-08-19 10:10
Article
S0003591
ID record
639227309946882470-7354
ID production
Temperature
Humidity
Test duration
15
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5583.80977 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1000.47136 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.691509105 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.673346585 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 685314.766 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.1357045 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29176062 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28504116 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.29033267 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79982085 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9933627 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0116857 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 12.0154897 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-08-19 10:10 1000.471 950 1050
2026-08-19 10:11 1000.451 950 1050
2026-08-19 10:12 1000.433 950 1050