Serial C00000724

Tests3
Approved2
Failures1

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-08-18 09:04 82903 639226406742062324-7499 S0003591 Fail View
2026-08-18 09:06 82904 639226407338566746-5740 S0003591 Pass View
2026-08-18 09:08 82907 639226408345316573-2152 S0003591 Pass View

Test 82903

Result
Fail
Date
2026-08-18 09:04
Article
S0003591
ID record
639226406742062324-7499
ID production
Temperature
Humidity
Test duration
15
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5622.8583 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1000.57746 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.710354653 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.690349295 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 690874.189 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.12385274 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29144328 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28477532 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.29000507 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.80130311 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9968126 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0120938 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 12.0139759 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-08-18 09:04 1000.577 950 1050
2026-08-18 09:06 1000.583 950 1050
2026-08-18 09:08 1000.539 950 1050