Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000577

Tests3
Approved2
Failures1

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-08-18 08:38 82874 639226390407231392-6514 S0003591 Pass View
2026-08-18 08:40 82876 639226392283043216-5203 S0003591 Fail View
2026-08-18 08:42 82879 639226392818591339-7978 S0003591 Pass View

Test 82876

Result
Fail
Date
2026-08-18 08:40
Article
S0003591
ID record
639226392283043216-5203
ID production
Temperature
Humidity
Test duration
15
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5585.21158 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 999.20572 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.708506433 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.688771175 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 693007.602 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.17008089 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.28847903 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28045338 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.26644865 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79701097 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9884959 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00883174 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.987615 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-08-18 08:38 999.214 950 1050
2026-08-18 08:40 999.206 950 1050
2026-08-18 08:42 999.155 950 1050