Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000549

Tests3
Approved2
Failures1

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-08-17 12:03 82739 639225649762361741-7484 S0003591 Pass View
2026-08-17 12:04 82740 639225650572741301-4296 S0003591 Fail View
2026-08-17 12:11 82741 639225654124138246-1132 S0003591 Pass View

Test 82740

Result
Fail
Date
2026-08-17 12:04
Article
S0003591
ID record
639225650572741301-4296
ID production
Temperature
Humidity
Test duration
15
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5589.47917 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1001.88001 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.700299239 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.925249796 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1444763.71 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.1830277 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.28703767 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.27906858 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.26489817 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.80097408 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.974897 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00833788 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 12.0131073 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-08-17 12:03 1001.914 950 1050
2026-08-17 12:04 1001.88 950 1050
2026-08-17 12:11 1001.879 950 1050