Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000554

Tests4
Approved2
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-08-17 11:53 82730 639225643346105627-6205 S0003591 Pass View
2026-08-17 11:54 82731 639225644244050801-5009 S0003591 Fail View
2026-08-17 11:55 82732 639225645002942630-8424 S0003591 Fail View
2026-08-17 11:57 82733 639225645736214123-3056 S0003591 Pass View

Test 82731

Result
Fail
Date
2026-08-17 11:54
Article
S0003591
ID record
639225644244050801-5009
ID production
Temperature
Humidity
Test duration
15
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5598.1025 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1001.48369 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.701484861 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.715245466 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1370322.68 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.22429371 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.28886438 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28131692 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.26509551 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79804748 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9502279 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00945018 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9973198 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (4)
DateValueMinMax
2026-08-17 11:53 1001.546 950 1050
2026-08-17 11:54 1001.484 950 1050
2026-08-17 11:55 1001.482 950 1050
2026-08-17 11:57 1001.474 950 1050