Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C01000584

Tests5
Approved2
Failures3

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-08-11 08:37 81319 639220342325691843-8129 S0003590 Fail View
2026-08-11 08:38 81320 639220342888104190-4418 S0003590 Fail View
2026-08-11 08:40 81321 639220344164868524-3708 S0003590 Fail View
2026-08-11 08:42 81322 639220345026274280-5858 S0003590 Pass View
2026-08-11 08:44 81323 639220346110681121-7962 S0003590 Pass View

Test 81321

Result
Fail
Date
2026-08-11 08:40
Article
S0003590
ID record
639220344164868524-3708
ID production
Temperature
Humidity
Test duration
30
Error ID
186
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5614.7865 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 997.550316 950 1050 Pass
1.2.1 Kontrola vypočítaného průměru z odporů: F1 (TP9/TP8) 0.633275 0 10 Pass
1.2.2 Kontrola vypočítaného průměru z odporů: F2 (TP10/TP11) 5.210381 0 10 Pass
1.2.3 Kontrola vypočítaného průměru z odporů: R3+R4 (TP10/TP8) 420192.267714 400000 460000 Pass
2.2.1 DMM: Měření napětí: +3V3 LDO OUTPUT (TP2/TP13) 3.28735801 3.1 3.4 Pass
2.2.2 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP15) 3.28138825 3.1 3.4 Pass
2.2.3 DMM: Měření napětí: +3V3 VDDA (TP2/TP24) 3.2868725 3.1 3.4 Pass
2.2.4 DMM: Měření napětí: +1V8 Audio (TP2/TP26) 1.80031283 1.74 1.86 Pass
2.3.1 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9708124 11.7 12.3 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (5)
DateValueMinMax
2026-08-11 08:37 997.499 950 1050
2026-08-11 08:38 997.526 950 1050
2026-08-11 08:40 997.55 950 1050
2026-08-11 08:42 997.94 950 1050
2026-08-11 08:44 997.964 950 1050