Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C01000330

Tests6
Approved3
Failures3

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-02-19 10:28 53481 639070936501776883-4198 S0003590 Fail View
2026-02-19 10:29 53482 639070937293700037-2540 S0003590 Pass View
2026-02-20 07:35 53662 639071696605411468-9238 S0003590 Fail View
2026-02-20 07:36 53663 639071697427275788-5362 S0003590 Pass View
2026-07-14 11:06 78423 639196239640203612-1052 S0003590 Fail View
2026-07-14 11:08 78424 639196240306668403-6245 S0003590 Pass View

Test 78423

Result
Fail
Date
2026-07-14 11:06
Article
S0003590
ID record
639196239640203612-1052
ID production
Temperature
Humidity
Test duration
30
Error ID
186
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5609.35626 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 999.35483 950 1050 Pass
1.2.1 Kontrola vypočítaného průměru z odporů: F1 (TP9/TP8) 0.574355 0 10 Pass
1.2.2 Kontrola vypočítaného průměru z odporů: F2 (TP10/TP11) 0.700504 0 10 Pass
1.2.3 Kontrola vypočítaného průměru z odporů: R3+R4 (TP10/TP8) 420493.584286 400000 460000 Pass
2.2.1 DMM: Měření napětí: +3V3 LDO OUTPUT (TP2/TP13) 3.28600393 3.1 3.4 Pass
2.2.2 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP15) 3.28028202 3.1 3.4 Pass
2.2.3 DMM: Měření napětí: +3V3 VDDA (TP2/TP24) 3.28581283 3.1 3.4 Pass
2.2.4 DMM: Měření napětí: +1V8 Audio (TP2/TP26) 1.80024411 1.74 1.86 Pass
2.3.1 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9372919 11.7 12.3 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (2)
DateValueMinMax
2026-07-14 11:06 999.355 950 1050
2026-07-14 11:08 999.345 950 1050