Serial lookup
Search a serial number to see its full test history.
Serial C01000287
Tests10
Approved2
Failures8
Tests over time
Test history
| Date | Measurement ID | ID record | Article | Result | |
|---|---|---|---|---|---|
| 2026-02-12 10:48 | 53158 | 639064900720654239-2435 | S0003590 | Fail | View |
| 2026-02-12 10:50 | 53159 | 639064901541424989-8059 | S0003590 | Fail | View |
| 2026-02-12 10:51 | 53160 | 639064902211119095-1767 | S0003590 | Fail | View |
| 2026-02-12 10:52 | 53161 | 639064902974854437-3867 | S0003590 | Fail | View |
| 2026-02-12 10:53 | 53162 | 639064903623026612-7732 | S0003590 | Fail | View |
| 2026-02-12 10:54 | 53163 | 639064904232514279-2244 | S0003590 | Fail | View |
| 2026-02-12 10:55 | 53164 | 639064904900695916-3696 | S0003590 | Fail | View |
| 2026-02-12 10:57 | 53165 | 639064905687165512-8436 | S0003590 | Fail | View |
| 2026-02-12 10:58 | 53166 | 639064906317577377-2575 | S0003590 | Pass | View |
| 2026-07-14 09:12 | 78379 | 639196171021361664-6106 | S0003590 | Pass | View |
Test 78379
- Result
- Pass
- Date
- 2026-07-14 09:12
- Article
- S0003590
- ID record
- 639196171021361664-6106
- ID production
- Temperature
- Humidity
- Test duration
- 69
- Error ID
- —
- Error text
- —
Measurement lines
| Step | Name | Value | Min | Max | Result |
|---|---|---|---|---|---|
| 1.1.2 | DMM: Měření odporu: R2 (TP5/TP6) | 1001.52125 | 950 | 1050 | Pass |
| 1.1.1 | DMM: Měření odporu: R1 (TP5/TP4) | 5600.32502 | 5320 | 5880 | Pass |
| 1.2.1 | Kontrola vypočítaného průměru z odporů: F1 (TP9/TP8) | 0.691516 | 0 | 10 | Pass |
| 1.2.2 | Kontrola vypočítaného průměru z odporů: F2 (TP10/TP11) | 0.810714 | 0 | 10 | Pass |
| 1.2.3 | Kontrola vypočítaného průměru z odporů: R3+R4 (TP10/TP8) | 420990.750857 | 400000 | 460000 | Pass |
| 2.2.1 | DMM: Měření napětí: +3V3 LDO OUTPUT (TP2/TP13) | 3.29986019 | 3.1 | 3.4 | Pass |
| 2.2.2 | DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP15) | 3.29470517 | 3.1 | 3.4 | Pass |
| 2.2.3 | DMM: Měření napětí: +3V3 VDDA (TP2/TP24) | 3.30031466 | 3.1 | 3.4 | Pass |
| 2.2.4 | DMM: Měření napětí: +1V8 Audio (TP2/TP26) | 1.7986676 | 1.74 | 1.86 | Pass |
| 2.3.1 | DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) | 12.0223889 | 11.7 | 12.3 | Pass |
| 3.2.1 | DMM: Měření frekvence: 1kHz (TP2/TP16) | 1000.00279 | 999.8 | 1000.2 | Pass |
| 3.3.2 | NZ3.2: Měření proudu: Kontrola odběru | 0.004992583 | < | 0.2 | Pass |
| 3.3.1 | NZ3.2: Měření proudu: Kontrola odběru | 0.005033847 | < | 0.02 | Pass |
| 3.4.1 | DMM: Měření odporu: (TP33/TP32) | 9.9E+37 | > | 10000000 | Pass |
| 3.4.2 | DMM: Měření odporu: (TP33/TP32) | 0.47980623 | < | 7 | Pass |
| 3.4.3 | DMM: Měření odporu: (TP35/TP34) | 9.9E+37 | > | 10000000 | Pass |
| 3.4.4 | DMM: Měření odporu: (TP35/TP34) | 0.544521668 | < | 7 | Pass |
| 3.4.5 | DMM: Měření odporu: (TP37/TP36) | 9.9E+37 | > | 10000000 | Pass |
| 3.4.6 | DMM: Měření odporu: (TP37/TP36) | 0.53034202 | < | 7 | Pass |
| 3.4.7 | DMM: Měření odporu: (TP39/TP40) | 9.9E+37 | > | 10000000 | Pass |
| 3.4.8 | DMM: Měření odporu: (TP39/TP40) | 116.104351 | 90 | 150 | Pass |
| 3.5.1 | TestSW: commPLM test | 0 | 0 | 0 | Pass |
| 3.5.2 | TestSW: commPLM test | 0 | 0 | 0 | Pass |
| 3.5.3 | TestSW: commPLM test | 0 | 0 | 0 | Pass |
Measurements over time
Hover for exact values · drag to zoom · double-click to reset
Readings (1)
| Date | Value | Min | Max |
|---|---|---|---|
| 2026-07-14 09:12 | 1001.521 | 950 | 1050 |