Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C01000253

Tests5
Approved2
Failures3

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-11 11:00 23152 639010475707502979-4375 S0003590 Fail View
2025-12-11 11:01 23153 639010476383630878-7497 S0003590 Pass View
2026-07-14 04:22 78206 639195997554200522-7219 S0003590 Fail View
2026-07-14 04:23 78207 639195998084084794-2345 S0003590 Fail View
2026-07-14 04:25 78208 639195998774794356-9009 S0003590 Pass View

Test 78207

Result
Fail
Date
2026-07-14 04:23
Article
S0003590
ID record
639195998084084794-2345
ID production
Temperature
Humidity
Test duration
30
Error ID
186
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5613.33456 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 999.538663 950 1050 Pass
1.2.1 Kontrola vypočítaného průměru z odporů: F1 (TP9/TP8) 2.94313 0 10 Pass
1.2.2 Kontrola vypočítaného průměru z odporů: F2 (TP10/TP11) 0.690737 0 10 Pass
1.2.3 Kontrola vypočítaného průměru z odporů: R3+R4 (TP10/TP8) 420331.519571 400000 460000 Pass
2.2.1 DMM: Měření napětí: +3V3 LDO OUTPUT (TP2/TP13) 3.2807616 3.1 3.4 Pass
2.2.2 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP15) 3.27430402 3.1 3.4 Pass
2.2.3 DMM: Měření napětí: +3V3 VDDA (TP2/TP24) 3.27999022 3.1 3.4 Pass
2.2.4 DMM: Měření napětí: +1V8 Audio (TP2/TP26) 1.79807071 1.74 1.86 Pass
2.3.1 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9461468 11.7 12.3 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-07-14 04:22 999.55 950 1050
2026-07-14 04:23 999.539 950 1050
2026-07-14 04:25 999.587 950 1050