Serial lookup
Search a serial number to see its full test history.
Serial C01000164
Tests3
Approved2
Failures1
Tests over time
Test history
| Date | Measurement ID | ID record | Article | Result | |
|---|---|---|---|---|---|
| 2025-12-11 14:11 | 23296 | 639010590167405762-6288 | S0003590 | Pass | View |
| 2026-07-13 08:05 | 77922 | 639195267189757958-4812 | S0003590 | Fail | View |
| 2026-07-13 08:07 | 77923 | 639195268051812252-6178 | S0003590 | Pass | View |
Test 77922
- Result
- Fail
- Date
- 2026-07-13 08:05
- Article
- S0003590
- ID record
- 639195267189757958-4812
- ID production
- Temperature
- Humidity
- Test duration
- 30
- Error ID
- 186
- Error text
- Vyhodnocení programu: [testSW]
Measurement lines
| Step | Name | Value | Min | Max | Result |
|---|---|---|---|---|---|
| 1.1.1 | DMM: Měření odporu: R1 (TP5/TP4) | 5635.31189 | 5320 | 5880 | Pass |
| 1.1.2 | DMM: Měření odporu: R2 (TP5/TP6) | 998.772546 | 950 | 1050 | Pass |
| 1.2.1 | Kontrola vypočítaného průměru z odporů: F1 (TP9/TP8) | 3.952536 | 0 | 10 | Pass |
| 1.2.2 | Kontrola vypočítaného průměru z odporů: F2 (TP10/TP11) | 1.823966 | 0 | 10 | Pass |
| 1.2.3 | Kontrola vypočítaného průměru z odporů: R3+R4 (TP10/TP8) | 420777.371857 | 400000 | 460000 | Pass |
| 2.2.1 | DMM: Měření napětí: +3V3 LDO OUTPUT (TP2/TP13) | 3.28578296 | 3.1 | 3.4 | Pass |
| 2.2.2 | DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP15) | 3.28011865 | 3.1 | 3.4 | Pass |
| 2.2.3 | DMM: Měření napětí: +3V3 VDDA (TP2/TP24) | 3.28572125 | 3.1 | 3.4 | Pass |
| 2.2.4 | DMM: Měření napětí: +1V8 Audio (TP2/TP26) | 1.79522966 | 1.74 | 1.86 | Pass |
| 2.3.1 | DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) | 11.9787735 | 11.7 | 12.3 | Pass |
Measurements over time
Hover for exact values · drag to zoom · double-click to reset
Readings (2)
| Date | Value | Min | Max |
|---|---|---|---|
| 2026-07-13 08:05 | 998.773 | 950 | 1050 |
| 2026-07-13 08:07 | 998.741 | 950 | 1050 |