Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C01000192

Tests5
Approved2
Failures3

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-11 14:35 23315 639010604721088779-8613 S0003590 Fail View
2025-12-11 14:36 23316 639010605301119528-2599 S0003590 Fail View
2025-12-11 14:37 23317 639010605943675255-4161 S0003590 Pass View
2026-07-01 06:11 76209 639184830314444486-2131 S0003590 Fail View
2026-07-01 06:12 76210 639184831007252212-3464 S0003590 Pass View

Test 76209

Result
Fail
Date
2026-07-01 06:11
Article
S0003590
ID record
639184830314444486-2131
ID production
Temperature
Humidity
Test duration
30
Error ID
186
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5592.20258 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 997.732009 950 1050 Pass
1.2.1 Kontrola vypočítaného průměru z odporů: F1 (TP9/TP8) 0.57087 0 10 Pass
1.2.2 Kontrola vypočítaného průměru z odporů: F2 (TP10/TP11) 1.139041 0 10 Pass
1.2.3 Kontrola vypočítaného průměru z odporů: R3+R4 (TP10/TP8) 420823.508143 400000 460000 Pass
2.2.1 DMM: Měření napětí: +3V3 LDO OUTPUT (TP2/TP13) 3.2783451 3.1 3.4 Pass
2.2.2 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP15) 3.27232885 3.1 3.4 Pass
2.2.3 DMM: Měření napětí: +3V3 VDDA (TP2/TP24) 3.27792995 3.1 3.4 Pass
2.2.4 DMM: Měření napětí: +1V8 Audio (TP2/TP26) 1.8012551 1.74 1.86 Pass
2.3.1 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9452078 11.7 12.3 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (2)
DateValueMinMax
2026-07-01 06:11 997.732 950 1050
2026-07-01 06:12 997.77 950 1050