Serial C00000444

Tests4
Approved1
Failures3

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-02-23 09:37 53734 639074361995034762-9388 S0003591 Fail View
2026-02-23 09:37 53735 639074362454293481-7725 S0003591 Fail View
2026-02-23 09:38 53736 639074363076240340-3995 S0003591 Fail View
2026-02-23 09:40 53737 639074363733324210-1716 S0003591 Pass View

Test 53736

Result
Fail
Date
2026-02-23 09:38
Article
S0003591
ID record
639074363076240340-3995
ID production
Temperature
Humidity
Test duration
25
Error ID
218
Error text
TestSW: test oscilátoru ON

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5631.7966 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1002.19425 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.391656598 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.572497146 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1398814.19 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.17385683 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.28749616 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.27998825 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.27801593 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79864311 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.984878 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.015696 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9616578 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (4)
DateValueMinMax
2026-02-23 09:37 1002.197 950 1050
2026-02-23 09:37 1002.156 950 1050
2026-02-23 09:38 1002.194 950 1050
2026-02-23 09:40 1002.11 950 1050