Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000471

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-02-23 09:06 53706 639074343603725898-6757 S0003591 Fail View
2026-02-23 09:07 53707 639074344085594290-7734 S0003591 Fail View
2026-02-23 09:08 53708 639074344673700838-2335 S0003591 Pass View

Test 53707

Result
Fail
Date
2026-02-23 09:07
Article
S0003591
ID record
639074344085594290-7734
ID production
Temperature
Humidity
Test duration
14
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5604.56849 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 998.400524 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.62287465 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.820204137 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1412360 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.18203287 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29326518 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28531313 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.28371523 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79681539 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9731966 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00924217 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9600124 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-02-23 09:06 998.396 950 1050
2026-02-23 09:07 998.401 950 1050
2026-02-23 09:08 998.373 950 1050