Serial C00000482

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-02-23 08:47 53691 639074332440108243-9243 S0003591 Fail View
2026-02-23 08:48 53692 639074332992156317-4245 S0003591 Fail View
2026-02-23 08:50 53693 639074333948880177-8516 S0003591 Pass View

Test 53692

Result
Fail
Date
2026-02-23 08:48
Article
S0003591
ID record
639074332992156317-4245
ID production
Temperature
Humidity
Test duration
36
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5607.31747 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 999.479307 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.606603066 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.790110704 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1346768.67 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.17124769 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29363085 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28642941 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.2848655 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79861721 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9861536 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00767802 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9438376 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-02-23 08:47 999.455 950 1050
2026-02-23 08:48 999.479 950 1050
2026-02-23 08:50 999.498 950 1050