Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000412

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-02-19 12:39 53520 639071015215936768-6673 S0003591 Fail View
2026-02-19 12:40 53521 639071015748182860-4776 S0003591 Fail View
2026-02-19 12:41 53522 639071016263484576-7138 S0003591 Pass View

Test 53521

Result
Fail
Date
2026-02-19 12:40
Article
S0003591
ID record
639071015748182860-4776
ID production
Temperature
Humidity
Test duration
26
Error ID
218
Error text
TestSW: test oscilátoru ON

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5602.91567 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 999.419179 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.637722324 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.826475069 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1365713.87 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.17006855 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29193698 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28499964 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.28285017 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79848773 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9832757 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0218479 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.970811 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-02-19 12:39 999.459 950 1050
2026-02-19 12:40 999.419 950 1050
2026-02-19 12:41 999.404 950 1050