Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C01R00002

Tests12
Approved3
Failures9

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-11 14:50 23328 639010614176010688-1632 S0003590 Fail View
2025-12-11 14:51 23329 639010614616220679-6748 S0003590 Fail View
2025-12-11 14:54 23331 639010615436057208-2096 S0003590 Fail View
2025-12-11 14:57 23332 639010617152904047-3909 S0003590 Fail View
2025-12-11 14:57 23333 639010618499781928-4824 S0003590 Fail View
2025-12-11 15:14 23335 639010627903341318-6008 S0003590 Fail View
2025-12-11 15:15 23336 639010628722768229-3112 S0003590 Pass View
2026-02-19 09:45 53463 639070910846388478-2622 S0003590 Fail View
2026-02-19 09:45 53464 639070911409859225-2707 S0003590 Fail View
2026-02-19 09:46 53465 639070911839824534-5303 S0003590 Fail View
2026-02-19 09:49 53466 639070912814778161-9516 S0003590 Pass View
2026-06-19 06:59 74831 639174490728470061-8437 S0003590 Pass View

Test 53465

Result
Fail
Date
2026-02-19 09:46
Article
S0003590
ID record
639070911839824534-5303
ID production
Temperature
Humidity
Test duration
31
Error ID
124
Error text
Kontrola vypočítaného průměru z odporů: F2 (TP10/TP11)

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5606.96545 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 998.481161 950 1050 Pass
1.2.1 Kontrola vypočítaného průměru z odporů: F1 (TP9/TP8) 1.101644 0 2 Pass
1.2.2 Kontrola vypočítaného průměru z odporů: F2 (TP10/TP11) 2.730139 0 2 Fail

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (1)
DateValueMinMax
2026-06-19 06:59 998.224 950 1050