Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000376

Tests4
Approved1
Failures3

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-02-16 09:55 53386 639068325213450584-8926 S0003591 Fail View
2026-02-16 09:56 53387 639068325822480231-8924 S0003591 Fail View
2026-02-16 09:57 53388 639068326397222672-9795 S0003591 Fail View
2026-02-16 09:59 53389 639068327128292577-5428 S0003591 Pass View

Test 53388

Result
Fail
Date
2026-02-16 09:57
Article
S0003591
ID record
639068326397222672-9795
ID production
Temperature
Humidity
Test duration
26
Error ID
218
Error text
TestSW: test oscilátoru ON

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5608.74297 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 999.029831 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.616096039 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.638862116 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1420166.54 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.15133553 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29451363 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28788211 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.29322244 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.80141438 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9935668 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0215788 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9646121 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (4)
DateValueMinMax
2026-02-16 09:55 999.04 950 1050
2026-02-16 09:56 999.041 950 1050
2026-02-16 09:57 999.03 950 1050
2026-02-16 09:59 999.026 950 1050