Serial C00000381

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2026-02-16 09:46 53378 639068319441765844-6555 S0003591 Fail View
2026-02-16 09:47 53379 639068320298637318-2779 S0003591 Fail View
2026-02-16 09:48 53380 639068320907827203-4533 S0003591 Pass View

Test 53379

Result
Fail
Date
2026-02-16 09:47
Article
S0003591
ID record
639068320298637318-2779
ID production
Temperature
Humidity
Test duration
15
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5592.58087 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1000.17153 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.622612336 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.631265095 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1427762.77 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.17780999 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29186229 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28434077 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.28246081 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79601089 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9704339 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0108241 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9261218 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2026-02-16 09:46 1000.212 950 1050
2026-02-16 09:47 1000.172 950 1050
2026-02-16 09:48 1000.148 950 1050