Serial lookup
Search a serial number to see its full test history.
Serial C01000440
Tests8
Approved2
Failures6
Tests over time
Test history
| Date | Measurement ID | ID record | Article | Result | |
|---|---|---|---|---|---|
| 2026-02-12 07:14 | 53011 | 639064772366950027-3109 | S0003590 | Fail | View |
| 2026-02-12 07:16 | 53012 | 639064773133021544-3139 | S0003590 | Fail | View |
| 2026-02-12 07:18 | 53013 | 639064774002308446-3945 | S0003590 | Fail | View |
| 2026-02-12 07:19 | 53014 | 639064775050083574-6166 | S0003590 | Fail | View |
| 2026-02-12 07:20 | 53015 | 639064775647728466-8608 | S0003590 | Fail | View |
| 2026-02-12 07:22 | 53016 | 639064776666746142-2075 | S0003590 | Fail | View |
| 2026-02-12 07:34 | 53024 | 639064783927555235-1050 | S0003590 | Pass | View |
| 2026-07-13 09:59 | 77994 | 639195334827975264-5997 | S0003590 | Pass | View |
Test 53012
- Result
- Fail
- Date
- 2026-02-12 07:16
- Article
- S0003590
- ID record
- 639064773133021544-3139
- ID production
- Temperature
- Humidity
- Test duration
- 77
- Error ID
- 270
- Error text
- TestSW: commPLM test
Measurement lines
| Step | Name | Value | Min | Max | Result |
|---|---|---|---|---|---|
| 1.1.2 | DMM: Měření odporu: R2 (TP5/TP6) | 1000.76598 | 950 | 1050 | Pass |
| 1.1.1 | DMM: Měření odporu: R1 (TP5/TP4) | 5616.57476 | 5320 | 5880 | Pass |
| 1.2.1 | Kontrola vypočítaného průměru z odporů: F1 (TP9/TP8) | 1.854037 | 0 | 2 | Pass |
| 1.2.2 | Kontrola vypočítaného průměru z odporů: F2 (TP10/TP11) | 0.802878 | 0 | 2 | Pass |
| 1.2.3 | Kontrola vypočítaného průměru z odporů: R3+R4 (TP10/TP8) | 420696.156714 | 400000 | 460000 | Pass |
| 2.2.1 | DMM: Měření napětí: +3V3 LDO OUTPUT (TP2/TP13) | 3.28861871 | 3.1 | 3.4 | Pass |
| 2.2.2 | DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP15) | 3.28221682 | 3.1 | 3.4 | Pass |
| 2.2.3 | DMM: Měření napětí: +3V3 VDDA (TP2/TP24) | 3.28827141 | 3.1 | 3.4 | Pass |
| 2.2.4 | DMM: Měření napětí: +1V8 Audio (TP2/TP26) | 1.79597093 | 1.74 | 1.86 | Pass |
| 2.3.1 | DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) | 11.9111408 | 11.7 | 12.3 | Pass |
| 3.2.1 | DMM: Měření frekvence: 1kHz (TP2/TP16) | 1000.00919 | 999.8 | 1000.2 | Pass |
| 3.3.1 | NZ3.2: Měření proudu: Kontrola odběru | 0.005036794 | < | 0.02 | Pass |
| 3.3.2 | NZ3.2: Měření proudu: Kontrola odběru | 0.005033847 | < | 0.2 | Pass |
| 3.4.1 | DMM: Měření odporu: (TP33/TP32) | 9.9E+37 | > | 10000000 | Pass |
| 3.4.2 | DMM: Měření odporu: (TP33/TP32) | 0.376451744 | < | 7 | Pass |
| 3.4.3 | DMM: Měření odporu: (TP35/TP34) | 9.9E+37 | > | 10000000 | Pass |
| 3.4.4 | DMM: Měření odporu: (TP35/TP34) | 2.48905071 | < | 7 | Pass |
| 3.4.5 | DMM: Měření odporu: (TP37/TP36) | 9.9E+37 | > | 10000000 | Pass |
| 3.4.6 | DMM: Měření odporu: (TP37/TP36) | 0.534203069 | < | 7 | Pass |
| 3.4.7 | DMM: Měření odporu: (TP39/TP40) | 9.9E+37 | > | 10000000 | Pass |
| 3.4.8 | DMM: Měření odporu: (TP39/TP40) | 125.426654 | 90 | 150 | Pass |
Measurements over time
Hover for exact values · drag to zoom · double-click to reset
Readings (1)
| Date | Value | Min | Max |
|---|---|---|---|
| 2026-07-13 09:59 | 1000.71 | 950 | 1050 |