Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial 001342222

Tests6
Approved1
Failures5

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2024-09-20 10:03 408 638624161747983393-2343 S0000608 Fail View
2024-09-20 10:03 409 638624162002647786-8195 S0000608 Fail View
2024-09-23 13:21 421 638626872630102339-7970 S0000608 Fail View
2024-09-23 13:24 422 638626874525481015-9329 S0000608 Fail View
2024-09-24 13:45 530 638627751500878569-3016 S0000608 Fail View
2024-09-24 14:00 531 638627759526223893-1834 S0000608 Pass View

Test 530

Result
Fail
Date
2024-09-24 13:45
Article
S0000608
ID record
638627751500878569-3016
ID production
47663
Temperature
26,71
Humidity
42,98
Test duration
8
Error ID
91
Error text
DMM: Kontrola hodnoty napětí (TP19/TP17,18)

Measurement lines

StepNameValueMinMaxResult
0.1 Uložení čísla výrobního příkazu 47663 Pass
0.2 Uložení ID obsluhy 0446 Pass
0.3 Uložení teploty okolí 26,71 Pass
0.4 Uložení vlhkosti okolí 42,98 Pass
0.5 Uložení Id testeru (uloženo na lokálním disku) Vez_Hokami_01-v01.01 Pass
0.6 Uložení id Artiklu (produktové číslo) S0000608 Pass
1.1 DMM: Kontrola hodnoty Odporu (TP2/TP3) 0,693229189 -1 1 Pass
1.2 DMM: Kontrola hodnoty Odporu (TP4/TP3) 5109,24462 5000 5200 Pass
1.3 DMM: Kontrola hodnoty Odporu (TP27/TP17,18) 10155,4611 9000 11000 Pass
2.1 DMM: Kontrola hodnoty Odporu (TP8/TP10) 1068,00269 1012 1120 Pass
4.1 NZ1: Odběr proudu ze zdroje (TP6/TP17,18) 0,00443561 0,0025 0,006 Pass
4.2 DMM: Kontrola hodnoty napětí (TP5/TP17,18) 11,9971023 11,9 12,1 Pass
4.3 DMM: Kontrola hodnoty napětí (TP19/TP17,18) 5,10163587 4,9 5,1 Fail

Measurements over time

No plottable values for this measurement.

Readings (1)
DateValueMinMax
2024-09-24 14:00