Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial 001855626

Tests7
Approved0
Failures7

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-05-26 09:13 15410 638838404027322848-7328 Fail View
2025-05-26 09:14 15411 638838404907482314-5406 Fail View
2025-05-26 09:15 15412 638838405395135477-1742 Fail View
2025-05-26 10:58 15413 638838467342543260-9448 Fail View
2025-05-26 13:22 15414 638838553725311653-6438 Fail View
2025-08-27 10:14 48000 638918792481797664-9975 Fail View
2026-02-02 07:09 60972 639056093645052555-9086 S0001604 Fail View

Test 48000

Result
Fail
Date
2025-08-27 10:14
Article
ID record
638918792481797664-9975
ID production
51291
Temperature
26.68
Humidity
33.55
Test duration
4
Error ID
057
Error text
NZ3: Měření proudu: (TP2/TP9)

Measurement lines

StepNameValueMinMaxResult
0.0 Dekódování a uložení načteného SN 001855626 Pass
0.1 Uložení čísla výrobního příkazu 51291 Pass
0.2 Uložení ID obsluhy 0736 Pass
0.3 Uložení teploty okolí 26.68 Pass
0.4 Uložení vlhkosti okolí 33.55 Pass
0.6 DMM: Status? 0000000000000000 xxxxxxxxx000000x Pass
0.7 NZ3: Status? 0000000000000000 xxxxxxxxx000000x Pass
0.8 EL: Status? 0000000000000000 xxxxxxxxx000000x Pass
1.1 DMM: Kontrola hodnoty Odporu (TP7/TP8) 100.616508 95 105 Pass
2.1 NZ3: Měření proudu: (TP2/TP9) 0.00827759 0.002 0.008 Fail

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (1)
DateValueMinMax
2026-02-02 07:09 1855626