Serial 001791472

Tests12
Approved1
Failures11

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2024-09-19 07:33 391 638623207428728363-5719 S0001055 Fail View
2024-09-19 07:35 392 638623209106098495-4337 S0001055 Fail View
2024-09-19 08:10 393 638623229949531348-4458 S0001055 Fail View
2024-12-05 14:54 5399 638690036072711655-9636 S0001055 Fail View
2024-12-05 15:21 5400 638690052492105046-5215 S0001055 Fail View
2025-09-05 11:27 25761 638926612215708654-3950 S0001055 Fail View
2025-11-06 17:29 40585 638980433396166126-5959 S0001055 Fail View
2026-07-09 14:48 77745 639192052730225671-9974 S0001055 Fail View
2026-07-09 14:49 77746 639192053380116589-3555 S0001055 Fail View
2026-08-11 04:41 81270 639220200196191802-4191 S0001055 Fail View
2026-08-11 04:51 81271 639220206265241579-4710 S0001055 Fail View
2026-09-23 05:29 86096 639257381296897102-3479 S0001055 Pass View

Test 40585

Result
Fail
Date
2025-11-06 17:29
Article
S0001055
ID record
638980433396166126-5959
ID production
49267
Temperature
26.4
Humidity
31.32
Test duration
55
Error ID
398
Error text
TestSW: [ADC_PA2] sense C

Measurement lines

StepNameValueMinMaxResult
0.0 Dekódování a uložení načteného SN 001791472 Pass
0.1 Uložení čísla výrobního příkazu 49267 Pass
0.2 Uložení ID obsluhy 1332 Pass
0.3 Uložení teploty okolí 26.4 Pass
0.4 Uložení vlhkosti okolí 31.32 Pass
0.5 Uložení Id testeru (uloženo na lokálním disku) Vez_Hokami_01-v01.01 Pass
0.6 Uložení id Artiklu (produktové číslo) S0001055 Pass
1.1 DMM: Kontrola hodnoty Odporu (TP6/TP4) 0.719134352 < 10 Pass
1.2 DMM: Kontrola hodnoty Odporu (TP6/TP5) 1000.41766 800 1200 Pass
3.1 NZ3.1: Měření napětí: Zkrat? 28.00851 > 27.9 Pass
3.2 NZ3.1: Měření proudu: Kontrola odběru 0.000814632 < 0.005 Pass
3.3 R11.12: Zapnutí (TP42/TP43) Pass
3.4 NZ3.1: Měření napětí: Zkrat? 28.00851 > 27.9 Pass
3.5 NZ3.1: Měření proudu: Kontrola odběru 0.02429511 < 0.04 Pass
4.1 DMM: Měření napětí: +BATT (TP49/TP44) 27.9693807 27.9 28.1 Pass
4.2 DMM: Měření napětí: +BATT (TP49/TP34) 27.9974235 27.9 28.1 Pass
4.3 DMM: Měření napětí: +12V (TP49/TP46) 11.8852828 11 12.2 Pass
4.4 DMM: Měření napětí: +5V (TP49/TP7) 5.00417124 4.9 5.1 Pass
4.5 DMM: Měření napětí: +3V3 (TP49/TP16) 3.29557893 3.2 3.4 Pass
4.6 DMM: Měření napětí: VDDA +3V3 (TP49/TP2) 3.29560642 3.2 3.4 Pass
4.7 DMM: Měření napětí: (TP49/TP8) 4.68430326 4.57 4.77 Pass
4.8 DMM: Měření napětí: (TP49/TP9) 4.68053933 4.57 4.77 Pass
4.9 DMM: Měření napětí: (TP49/TP10) 4.68108884 4.57 4.77 Pass
5.1 DMM: Měření napětí: (TP49/TP46) 0.516781692 < 1 Pass
5.2 DMM: Měření napětí: (TP49/TP46) 11.793183 > 11.5 Pass
6.1 R9.12: Vypnutí (TP49/TP28) Pass
7.1 NZ3: Měření napětí: Zkrat? 28.00851 > 27.9 Pass
7.2 NZ3: Měření proudu: Kontrola odběru 0.02536482 < 0.06 Pass
7.3 R0.6: Zapnutí Ext. Programátoru Pass
8.1 TestSw: Ověření komunikace [TYP:AA NAME:eDrive NAME2:TestSW HW:Kolecko v4 SW: Pass
9.1 TestSW: [CHG_SIG (PC14)] = log1? 16384 16384 16384 Pass
9.2 EL1.1: Ověření proudu na zátěži -> Current? 2.108409 1.6 2.8 Pass
9.3 DMM: Měření napětí: Vstup U14 (TP39/TP32) 29.5629573 29 30.1 Pass
9.4 DMM: Měření napětí: Vstup U14 (TP39/TP33) 11.2845791 10.8 11.5 Pass
9.5 DMM: Měření napětí: Vstup U14 (TP39/TP40) 1.61057585 1.5 1.8 Pass
10.1 NZ3.1: Měření napětí: Zkrat? 28.00851 > 27.9 Pass
10.2 NZ3.1: Měření proudu: Kontrola odběru 0.02547385 < 0.06 Pass
10.3 TestSW: [ADC_PA03] 3843 3667 3937 Pass
12.1 TestSW: [PA15] - napájení sense ABC Pass
12.2 TestSW: [HALL_A (PB4)] = log0? 0 0 0 Pass
12.3 TestSW: [HALL_A (PB4)] = log1? 16 16 16 Pass
12.4 TestSW: [HALL_B (PB5)] = log0? 0 0 0 Pass
12.5 TestSW: [HALL_B (PB5)] = log1? 32 32 32 Pass
12.6 TestSW: [HALL_C (PB0)] = log0? 0 0 0 Pass
12.7 TestSW: [HALL_C (PB0)] = log1? 1 1 1 Pass
13.1 TestSW: [CON_ID (PC13)] = log1? 8192 8192 8192 Pass
13.10 TestSW: [BTN_FFWD][ADC_PA6]? 4085 4000 4096 Pass
13.11 TestSW: [BTN_FFWD][ADC_PA6]? 2026 1900 2100 Pass
13.12 TestSW: [BTN_STOP][ADC_PA5]? 4086 4000 4096 Pass
13.13 TestSW: [BTN_STOP][ADC_PA5]? 2025 1900 2100 Pass
13.2 TestSW: [CON_ID (PC13)] = log0? 0 0 0 Pass
13.3 TestSW: [BTN_DRV][PA11] na log1 Pass
13.4 TestSW: [BTN_FWD][ADC_PB1]? 4085 4000 4096 Pass
13.5 TestSW: [BTN_FWD][ADC_PB1]? 2012 1900 2100 Pass
13.6 TestSW: [BTN_REV][ADC_PA4]? 4086 4000 4096 Pass
13.7 TestSW: [BTN_REV][ADC_PA4]? 2025 1900 2100 Pass
13.8 TestSW: [BTN_ULCK][ADC_PA7]? 4086 4000 4096 Pass
13.9 TestSW: [BTN_ULCK][ADC_PA7]? 2023 1900 2100 Pass
14.1 TestSW: [LED1][PB2] na log1 Pass
14.10 DMM: Měření napětí: LED3 (TP49/TP14) 3.29187766 > 3 Pass
14.11 TestSW: [LED3][PB11] na log0 Pass
14.12 DMM: Měření napětí: LED3 (TP49/TP14) 0.0034541386 < 0.1 Pass
14.13 TestSW: [LED4][PB12] na log1 Pass
14.14 DMM: Měření napětí: LED4 (TP49/TP12) 3.29317802 > 3 Pass
14.15 TestSW: [LED4][PB12] na log0 Pass
14.16 DMM: Měření napětí: LED4 (TP49/TP12) 0.00344591128 < 0.1 Pass
14.2 DMM: Měření napětí: LED1 (TP49/TP13) 3.29160652 > 3 Pass
14.3 TestSW: [LED1][PB2] na log0 Pass
14.4 DMM: Měření napětí: LED1 (TP49/TP13) 0.00357964967 < 0.1 Pass
14.5 TestSW: [LED2][PB10] na log1 Pass
14.6 DMM: Měření napětí: LED1 (TP49/TP11) 3.29179906 > 3 Pass
14.7 TestSW: [LED2][PB10] na log0 Pass
14.8 DMM: Měření napětí: LED1 (TP49/TP11) 0.0034709296 < 0.1 Pass
14.9 TestSW: [LED3][PB11] na log1 Pass
15.1 TestSW: ntc 55 49 90 Pass
16.1 DMM: Měření frekvence (TP49/TP55) 19995.1272 19600 20400 Pass
16.2 DMM: Měření frekvence (TP49/TP56) 19995.1287 19600 20400 Pass
16.3 DMM: Měření napětí (TP49/TP52) 14.0065101 13.3 14.7 Pass
17.1 DMM: Měření frekvence (TP49/TP61) 19995.123 19600 20400 Pass
17.2 DMM: Měření frekvence (TP49/TP62) 19995.1268 19600 20400 Pass
17.3 DMM: Měření napětí (TP49/TP58) 14.0070274 13.3 14.7 Pass
18.1 DMM: Měření frekvence (TP49/TP67) 19995.1211 19600 20400 Pass
18.2 DMM: Měření frekvence (TP49/TP68) 19995.1254 19600 20400 Pass
18.3 DMM: Měření napětí (TP49/TP58) 14.0109909 13.3 14.7 Pass
19.1 DMM: Měření napětí: +5V (TP49/TP7) 0.00173344795 < 0.5 Pass
20.0.1 TestSW: [ADC_PA0] sense A do var VA = 2276 Pass
20.0.2 TestSW: [ADC_PA1] sense B do var VB = 2273 Pass
20.0.3 TestSW: [ADC_PA2] sense C do var VC = 2276 Pass
20.0.5 Analysis: VA VA[2276] 2210 2340 Pass
20.0.6 Analysis: VB VB[2273] 2210 2340 Pass
20.0.7 Analysis: VC VC[2276] 2210 2340 Pass
20.1 NZ3.1: Měření napětí: Zkrat? 28.00851 > 27.9 Pass
20.1.1 EL1.2: Ověření proudu na zátěži -> Current? 4.989653 4.7 5.3 Pass
20.1.2 TestSW: [ADC_PA0] sense A 2448 2441[VA] 2541[VA] Pass
20.1.3 TestSW: [ADC_PA1] sense B 2104 2008[VB] 2108[VB] Pass
20.1.4 TestSW: [ADC_PA2] sense C 2276 2210 2340 Pass
20.2 NZ3.1: Měření proudu: Kontrola odběru 0.03788891 < 0.06 Pass
20.2.1 EL1.2: Ověření proudu na zátěži -> Current? 4.989779 4.7 5.3 Pass
20.2.2 TestSW: [ADC_PA0] sense A 2104 2011[VA] 2111[VA] Pass
20.2.3 TestSW: [ADC_PA1] sense B 2441 2438[VB] 2538[VB] Pass
20.2.4 TestSW: [ADC_PA2] sense C 2275 2210 2340 Pass
20.3.1 EL1.2: Ověření proudu na zátěži -> Current? 4.989686 4.7 5.3 Pass
20.3.2 TestSW: [ADC_PA0] sense A 2449 2441[VA] 2541[VA] Pass
20.3.3 TestSW: [ADC_PA1] sense B 2276 2210 2340 Pass
20.3.4 TestSW: [ADC_PA2] sense C 2110 2011[VC] 2111[VC] Pass
20.4.1 EL1.2: Ověření proudu na zátěži -> Current? 4.989855 4.7 5.3 Pass
20.4.2 TestSW: [ADC_PA0] sense A 2104 2011[VA] 2111[VA] Pass
20.4.3 TestSW: [ADC_PA1] sense B 2271 2210 2340 Pass
20.4.4 TestSW: [ADC_PA2] sense C 2440 2441[VC] 2541[VC] Fail

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (5)
DateValueMinMax
2026-07-09 14:48 1.022 10
2026-07-09 14:49 0.976 10
2026-08-11 04:41 0.771 10
2026-08-11 04:51 0.753 10
2026-09-23 05:29 0.811 10