| Date | Measurement ID | ID record | Article | Result | |
|---|---|---|---|---|---|
| 2024-09-09 07:11 | 266 | 638614554204821387-8479 | S0001055 | Pass | View |
| 2024-11-06 13:24 | 3077 | 638664926786709026-2808 | S0001055 | Fail | View |
| 2024-11-06 13:25 | 3078 | 638664927201711081-8662 | S0001055 | Fail | View |
| 2024-11-06 13:27 | 3079 | 638664927923652868-5737 | S0001055 | Pass | View |
| 2024-11-06 14:14 | 3080 | 638664955869338169-9816 | S0001055 | Pass | View |
| 2024-11-06 14:23 | 3081 | 638664957381735489-5047 | S0001055 | Pass | View |
| 2024-11-06 14:30 | 3082 | 638664964722500856-4415 | S0001055 | Pass | View |
| 2024-11-06 14:37 | 3083 | 638664969203024608-8496 | S0001055 | Pass | View |
| 2024-11-06 14:56 | 3084 | 638664980876475571-9000 | S0001055 | Pass | View |
| 2024-11-06 15:01 | 3085 | 638664984096327532-5332 | S0001055 | Pass | View |
| 2024-11-06 15:03 | 3086 | 638664986196528986-2405 | S0001055 | Fail | View |
| 2024-11-06 15:05 | 3087 | 638664986388451539-9375 | S0001055 | Pass | View |
Test 3082
- Result
- Pass
- Date
- 2024-11-06 14:30
- Article
- S0001055
- ID record
- 638664964722500856-4415
- ID production
- 44607
- Temperature
- 25,09
- Humidity
- 36,54
- Test duration
- 155
- Error ID
- —
- Error text
- —
Measurement lines
| Step | Name | Value | Min | Max | Result |
|---|---|---|---|---|---|
| 0.3 | Uložení teploty okolí | 25,09 | Pass | ||
| 0.4 | Uložení vlhkosti okolí | 36,54 | Pass | ||
| 0.5 | Uložení Id testeru (uloženo na lokálním disku) | Linet_01-v1.1 | Pass | ||
| 0.6 | Uložení id Artiklu (produktové číslo) | S0001055 | Pass | ||
| 1.1 | DMM: Kontrola hodnoty Odporu (TP6/TP4) | 0,41115521 | < | 10 | Pass |
| 1.2 | DMM: Kontrola hodnoty Odporu (TP6/TP5) | 1002,22084 | 800 | 1200 | Pass |
| 3.1 | NZ3.1: Měření napětí: Zkrat? | 28,00145 | > | 27,9 | Pass |
| 3.2 | NZ3.1: Měření proudu: Kontrola odběru | 0,0003804021 | < | 0,005 | Pass |
| 3.3 | R11.12: Zapnutí (TP42/TP43) | Pass | |||
| 3.4 | NZ3.1: Měření napětí: Zkrat? | 28,00145 | > | 27,9 | Pass |
| 3.5 | NZ3.1: Měření proudu: Kontrola odběru | 0,02384769 | < | 0,04 | Pass |
| 4.1 | DMM: Měření napětí: +BATT (TP49/TP44) | 27,9650673 | 27,9 | 28,1 | Pass |
| 4.2 | DMM: Měření napětí: +BATT (TP49/TP34) | 27,9961904 | 27,9 | 28,1 | Pass |
| 4.3 | DMM: Měření napětí: +12V (TP49/TP46) | 11,8134883 | 11 | 12,2 | Pass |
| 4.4 | DMM: Měření napětí: +5V (TP49/TP7) | 4,99436153 | 4,9 | 5,1 | Pass |
| 4.5 | DMM: Měření napětí: +3V3 (TP49/TP16) | 3,29066232 | 3,2 | 3,4 | Pass |
| 4.6 | DMM: Měření napětí: VDDA +3V3 (TP49/TP2) | 3,29066688 | 3,2 | 3,4 | Pass |
| 4.7 | DMM: Měření napětí: (TP49/TP8) | 4,67443724 | 4,57 | 4,77 | Pass |
| 4.8 | DMM: Měření napětí: (TP49/TP9) | 4,67124489 | 4,57 | 4,77 | Pass |
| 4.9 | DMM: Měření napětí: (TP49/TP10) | 4,67247536 | 4,57 | 4,77 | Pass |
| 5.1 | DMM: Měření napětí: (TP49/TP46) | 0,496916272 | < | 1 | Pass |
| 5.2 | DMM: Měření napětí: (TP49/TP46) | 11,7379318 | > | 11,5 | Pass |
| 6.1 | R9.12: Vypnutí (TP49/TP28) | Pass | |||
| 7.1 | NZ3: Měření napětí: Zkrat? | 28,00145 | > | 27,9 | Pass |
| 7.2 | NZ3: Měření proudu: Kontrola odběru | 0,0113109 | < | 0,06 | Pass |
| 7.3 | R0.6: Zapnutí Ext. Programátoru | Pass | |||
| 8.1 | TestSw: Ověření komunikace | [TYP:AA NAME:eDrive NAME2:TestSW HW:Kolecko v4 SW: | Pass | ||
| 9.1 | TestSW: [CHG_SIG (PC14)] = log1? | 16384 | 16384 | 16384 | Pass |
| 9.2 | EL1.1: Ověření proudu na zátěži -> Current? | 1,999756 | 1,6 | 2,8 | Pass |
| 9.3 | DMM: Měření napětí: Vstup U14 (TP39/TP32) | 29,7072301 | 29 | 30,1 | Pass |
| 9.4 | DMM: Měření napětí: Vstup U14 (TP39/TP33) | 11,2563573 | 10,8 | 11,5 | Pass |
| 9.5 | DMM: Měření napětí: Vstup U14 (TP39/TP40) | 1,57442894 | 1,5 | 1,8 | Pass |
| 10.1 | NZ3.1: Měření napětí: Zkrat? | 28,00345 | > | 27,9 | Pass |
| 10.2 | NZ3.1: Měření proudu: Kontrola odběru | 0,02572614 | < | 0,06 | Pass |
| 10.3 | TestSW: [ADC_PA03] | 3869 | 3667 | 3937 | Pass |
| 12.1 | TestSW: [PA15] - napájení sense ABC | Pass | |||
| 12.2 | TestSW: [HALL_A (PB4)] = log0? | 0 | 0 | 0 | Pass |
| 12.3 | TestSW: [HALL_A (PB4)] = log1? | 16 | 16 | 16 | Pass |
| 12.4 | TestSW: [HALL_B (PB5)] = log0? | 0 | 0 | 0 | Pass |
| 12.5 | TestSW: [HALL_B (PB5)] = log1? | 32 | 32 | 32 | Pass |
| 12.6 | TestSW: [HALL_C (PB0)] = log0? | 0 | 0 | 0 | Pass |
| 12.7 | TestSW: [HALL_C (PB0)] = log1? | 1 | 1 | 1 | Pass |
| 13.1 | TestSW: [CON_ID (PC13)] = log1? | 8192 | 8192 | 8192 | Pass |
| 13.10 | TestSW: [BTN_FFWD][ADC_PA6]? | 4088 | 4000 | 4096 | Pass |
| 13.11 | TestSW: [BTN_FFWD][ADC_PA6]? | 2028 | 1900 | 2100 | Pass |
| 13.12 | TestSW: [BTN_STOP][ADC_PA5]? | 4088 | 4000 | 4096 | Pass |
| 13.13 | TestSW: [BTN_STOP][ADC_PA5]? | 2028 | 1900 | 2100 | Pass |
| 13.2 | TestSW: [CON_ID (PC13)] = log0? | 0 | 0 | 0 | Pass |
| 13.3 | TestSW: [BTN_DRV][PA11] na log1 | Pass | |||
| 13.4 | TestSW: [BTN_FWD][ADC_PB1]? | 4088 | 4000 | 4096 | Pass |
| 13.5 | TestSW: [BTN_FWD][ADC_PB1]? | 2012 | 1900 | 2100 | Pass |
| 13.6 | TestSW: [BTN_REV][ADC_PA4]? | 4088 | 4000 | 4096 | Pass |
| 13.7 | TestSW: [BTN_REV][ADC_PA4]? | 2030 | 1900 | 2100 | Pass |
| 13.8 | TestSW: [BTN_ULCK][ADC_PA7]? | 4088 | 4000 | 4096 | Pass |
| 13.9 | TestSW: [BTN_ULCK][ADC_PA7]? | 2029 | 1900 | 2100 | Pass |
| 14.1 | TestSW: [LED1][PB2] na log1 | Pass | |||
| 14.10 | DMM: Měření napětí: LED3 (TP49/TP14) | 3,28651682 | > | 3 | Pass |
| 14.11 | TestSW: [LED3][PB11] na log0 | Pass | |||
| 14.12 | DMM: Měření napětí: LED3 (TP49/TP14) | 0,00356781428 | < | 0,1 | Pass |
| 14.13 | TestSW: [LED4][PB12] na log1 | Pass | |||
| 14.14 | DMM: Měření napětí: LED4 (TP49/TP12) | 3,2878409 | > | 3 | Pass |
| 14.15 | TestSW: [LED4][PB12] na log0 | Pass | |||
| 14.16 | DMM: Měření napětí: LED4 (TP49/TP12) | 0,00355168698 | < | 0,1 | Pass |
| 14.2 | DMM: Měření napětí: LED1 (TP49/TP13) | 3,28621759 | > | 3 | Pass |
| 14.3 | TestSW: [LED1][PB2] na log0 | Pass | |||
| 14.4 | DMM: Měření napětí: LED1 (TP49/TP13) | 0,00372081712 | < | 0,1 | Pass |
| 14.5 | TestSW: [LED2][PB10] na log1 | Pass | |||
| 14.6 | DMM: Měření napětí: LED1 (TP49/TP11) | 3,28644475 | > | 3 | Pass |
| 14.7 | TestSW: [LED2][PB10] na log0 | Pass | |||
| 14.8 | DMM: Měření napětí: LED1 (TP49/TP11) | 0,00359026389 | < | 0,1 | Pass |
| 14.9 | TestSW: [LED3][PB11] na log1 | Pass | |||
| 15.1 | TestSW: ntc | 61 | 49 | 90 | Pass |
| 16.1 | DMM: Měření frekvence (TP49/TP55) | 19994,9349 | 19600 | 20400 | Pass |
| 16.2 | DMM: Měření frekvence (TP49/TP56) | 19994,9349 | 19600 | 20400 | Pass |
| 16.3 | DMM: Měření napětí (TP49/TP52) | 13,9986707 | 13,3 | 14,7 | Pass |
| 17.1 | DMM: Měření frekvence (TP49/TP61) | 19994,9295 | 19600 | 20400 | Pass |
| 17.2 | DMM: Měření frekvence (TP49/TP62) | 19994,933 | 19600 | 20400 | Pass |
| 17.3 | DMM: Měření napětí (TP49/TP58) | 14,0016011 | 13,3 | 14,7 | Pass |
| 18.1 | DMM: Měření frekvence (TP49/TP67) | 19994,9274 | 19600 | 20400 | Pass |
| 18.2 | DMM: Měření frekvence (TP49/TP68) | 19994,9312 | 19600 | 20400 | Pass |
| 18.3 | DMM: Měření napětí (TP49/TP58) | 13,9970808 | 13,3 | 14,7 | Pass |
| 19.1 | DMM: Měření napětí: +5V (TP49/TP7) | 0,00107533902 | < | 0,5 | Pass |
| 20.0.1 | TestSW: [ADC_PA0] sense A do var | VA = 2277 | Pass | ||
| 20.0.2 | TestSW: [ADC_PA1] sense B do var | VB = 2271 | Pass | ||
| 20.0.3 | TestSW: [ADC_PA2] sense C do var | VC = 2279 | Pass | ||
| 20.0.5 | Analysis: VA | VA[2277] | 2210 | 2340 | Pass |
| 20.0.6 | Analysis: VB | VB[2271] | 2210 | 2340 | Pass |
| 20.0.7 | Analysis: VC | VC[2279] | 2210 | 2340 | Pass |
| 20.1 | NZ3.1: Měření napětí: Zkrat? | 28,00145 | > | 27,9 | Pass |
| 20.1.1 | EL1.2: Ověření proudu na zátěži -> Current? | 4,996858 | 4,7 | 5,3 | Pass |
| 20.1.2 | TestSW: [ADC_PA0] sense A | 2451 | 2442[VA] | 2542[VA] | Pass |
| 20.1.3 | TestSW: [ADC_PA1] sense B | 2099 | 2006[VB] | 2106[VB] | Pass |
| 20.1.4 | TestSW: [ADC_PA2] sense C | 2279 | 2210 | 2340 | Pass |
| 20.2 | NZ3.1: Měření proudu: Kontrola odběru | 0,03824518 | < | 0,06 | Pass |
| 20.2.1 | EL1.2: Ověření proudu na zátěži -> Current? | 4,996889 | 4,7 | 5,3 | Pass |
| 20.2.2 | TestSW: [ADC_PA0] sense A | 2103 | 2012[VA] | 2112[VA] | Pass |
| 20.2.3 | TestSW: [ADC_PA1] sense B | 2443 | 2436[VB] | 2536[VB] | Pass |
| 20.2.4 | TestSW: [ADC_PA2] sense C | 2278 | 2210 | 2340 | Pass |
| 20.3.1 | EL1.2: Ověření proudu na zátěži -> Current? | 4,996804 | 4,7 | 5,3 | Pass |
| 20.3.2 | TestSW: [ADC_PA0] sense A | 2451 | 2442[VA] | 2542[VA] | Pass |
| 20.3.3 | TestSW: [ADC_PA1] sense B | 2274 | 2210 | 2340 | Pass |
| 20.3.4 | TestSW: [ADC_PA2] sense C | 2103 | 2014[VC] | 2114[VC] | Pass |
| 20.4.1 | EL1.2: Ověření proudu na zátěži -> Current? | 4,996899 | 4,7 | 5,3 | Pass |
| 20.4.2 | TestSW: [ADC_PA0] sense A | 2102 | 2012[VA] | 2112[VA] | Pass |
| 20.4.3 | TestSW: [ADC_PA1] sense B | 2269 | 2210 | 2340 | Pass |
| 20.4.4 | TestSW: [ADC_PA2] sense C | 2453 | 2444[VC] | 2544[VC] | Pass |
| 20.5.1 | EL1.2: Ověření proudu na zátěži -> Current? | 4,996763 | 4,7 | 5,3 | Pass |
| 20.5.2 | TestSW: [ADC_PA0] sense A | 2277 | 2210 | 2340 | Pass |
| 20.5.3 | TestSW: [ADC_PA1] sense B | 2446 | 2436[VB] | 2536[VB] | Pass |
| 20.5.4 | TestSW: [ADC_PA2] sense C | 2103 | 2014[VC] | 2114[VC] | Pass |
| 20.6.1 | EL1.2: Ověření proudu na zátěži -> Current? | 4,996812 | 4,7 | 5,3 | Pass |
| 20.6.2 | TestSW: [ADC_PA0] sense A | 2277 | 2210 | 2340 | Pass |
| 20.6.3 | TestSW: [ADC_PA1] sense B | 2096 | 2006[VB] | 2106[VB] | Pass |
| 20.6.4 | TestSW: [ADC_PA2] sense C | 2453 | 2444[VC] | 2544[VC] | Pass |
| 23.1 | DMM: měření RMS(TP49/TP67) | 0,473624727 | 0,1 | 1,6 | Pass |
Measurements over time
Hover for exact values · drag to zoom · double-click to reset
Readings (12)
| Date | Value | Min | Max |
|---|---|---|---|
| 2024-09-09 07:11 | 28.34 | — | — |
| 2024-11-06 13:24 | 24.97 | — | — |
| 2024-11-06 13:25 | 24.97 | — | — |
| 2024-11-06 13:27 | 25.04 | — | — |
| 2024-11-06 14:14 | 25.1 | — | — |
| 2024-11-06 14:23 | 25.1 | — | — |
| 2024-11-06 14:30 | 25.09 | — | — |
| 2024-11-06 14:37 | 25.08 | — | — |
| 2024-11-06 14:56 | 25.24 | — | — |
| 2024-11-06 15:01 | 25.35 | — | — |
| 2024-11-06 15:03 | 25.35 | — | — |
| 2024-11-06 15:05 | 25.35 | — | — |