Serial lookup
Search a serial number to see its full test history.
Serial 001487991
Tests3
Approved1
Failures2
Tests over time
Test history
| Date | Measurement ID | ID record | Article | Result | |
|---|---|---|---|---|---|
| 2025-10-14 11:01 | 27756 | 638960292366352903-2450 | S0000174 | Fail | View |
| 2026-01-19 08:42 | 58383 | 639044053290593774-6171 | S0000174 | Fail | View |
| 2026-02-02 06:42 | 60713 | 639056077123812185-9786 | S0000174 | Pass | View |
Test 27756
- Result
- Fail
- Date
- 2025-10-14 11:01
- Article
- S0000174
- ID record
- 638960292366352903-2450
- ID production
- 53731
- Temperature
- 26.96
- Humidity
- 34.7
- Test duration
- 30
- Error ID
- 338
- Error text
- Analysis: hodnota na max
Measurement lines
| Step | Name | Value | Min | Max | Result |
|---|---|---|---|---|---|
| 0.0 | Dekódování a uložení načteného SN | 001487991 | Pass | ||
| 0.1 | Uložení čísla výrobního příkazu | 53731 | Pass | ||
| 0.2 | Uložení ID obsluhy | 1950 | Pass | ||
| 0.3 | Uložení teploty okolí | 26.96 | Pass | ||
| 0.4 | Uložení vlhkosti okolí | 34.7 | Pass | ||
| 0.5 | Uložení Id testeru (uloženo na lokálním disku) | Vez_Hokami_01-v01.01 | Pass | ||
| 0.6 | Uložení id Artiklu (produktové číslo) | S0000174 | Pass | ||
| 1.1 | DMM: Kontrola hodnoty Odporu (TP2/TP3) | 999.194573 | 900 | 1100 | Pass |
| 1.2 | DMM: Kontrola hodnoty Odporu (TP4/TP3) | 999.897485 | 900 | 1100 | Pass |
| 1.3 | DMM: Kontrola hodnoty Odporu (TP48/TP25,26) | 999.680353 | 900 | 1100 | Pass |
| 1.4 | DMM: Kontrola hodnoty Odporu (TP13/TP25,26) | 100361.172 | 90000 | 110000 | Pass |
| 2.1 | DMM: Kontrola hodnoty Odporu (TP36/TP37) | 1065.26703 | 1012 | 1120 | Pass |
| 2.2 | DMM: Kontrola hodnoty Odporu (TP40/TP41) | 1037.73316 | 1012 | 1120 | Pass |
| 4.0 | DMM: Kontrola hodnoty napětí (TP22/TP25,26) | -0.722964643 | -0.75 | -0.65 | Pass |
| 4.1 | NZ1: Odběr proudu ze zdroje (TP25/TP25,26) | 0.009986895 | 0.009 | 1 | Pass |
| 5.1 | NZ1: Odběr proudu ze zdroje (TP22/TP25,26) | 0.0308532 | 0.025 | 0.035 | Pass |
| 5.2 | DMM: Kontrola hodnoty napětí (TP49/TP25,26) | 8.7719827 | 8.7 | 9 | Pass |
| 5.3 | DMM: Kontrola hodnoty napětí (TP17/TP25,26) | 5.0428698 | 4.9 | 5.1 | Pass |
| 5.4 | DMM: Kontrola hodnoty napětí (TP16/TP25,26) | 5.00364168 | 4.9 | 5.1 | Pass |
| 5.5 | DMM: Kontrola hodnoty napětí (TP29/TP25,26) | 5.00368084 | 4.9 | 5.1 | Pass |
| 5.6 | DMM: Kontrola hodnoty napětí (TP14/TP25,26) | 3.29796774 | 3.2 | 3.4 | Pass |
| 5.7 | DMM: Kontrola hodnoty napětí (TP5/TP25,26) | 3.29780217 | 3.2 | 3.4 | Pass |
| 5.8 | DMM: Kontrola hodnoty napětí (TP24/TP25,26) | 8.70052743 | 7.5 | 9 | Pass |
| 5.9 | DMM: Kontrola hodnoty napětí (TP20/TP25,26) | 0.233353182 | 0 | 0.65 | Pass |
| 6.1 | NZ1: Odběr proudu ze zdroje (TP25,26/TP19) | 2.723677E-05 | -0.0001 | 0.0001 | Pass |
| 6.2 | DMM: Kontrola hodnoty napětí (TP20/TP25,26) | 0.363038096 | -0.2 | 0.5 | Pass |
| 6.3 | NZ1: Odběr proudu ze zdroje (TP25,26/TP19) | 0.01994655 | 0.019 | 1 | Pass |
| 7.1 | NZ1: Odběr proudu ze zdroje (TP19/TP25,26) | 5.989139E-05 | 0 | 0.0001 | Pass |
| 7.2 | DMM: Kontrola hodnoty napětí (TP20/TP25,26) | 6.0025378 | 5.95 | 7 | Pass |
| 7.3 | DMM: Kontrola hodnoty napětí (TP50/TP25,26) | 3.30926797 | 3.2 | 3.4 | Pass |
| 7.4 | DMM: Kontrola hodnoty napětí (TP15/TP25,26) | 0.143117484 | 0 | 0.5 | Pass |
| 7.5 | NZ1: Odběr proudu ze zdroje (TP19/TP25,26) | 0.0006150199 | 0.00057 | 0.00063 | Pass |
| 7.6 | NZ2: Odběr proudu ze zdroje (TP42/TP25,26) | 9.561545E-05 | -0.001 | 0.0002 | Pass |
| 7.7 | NZ1: Odběr proudu ze zdroje (TP19/TP25,26) | 2.723677E-05 | -0.001 | 0.0001 | Pass |
| 7.8 | DMM: Kontrola hodnoty napětí (TP38/TP25,26) | 8.95304068 | 8.7 | 9 | Pass |
| 9.1 | NZ1: Odběr proudu ze zdroje (TP22/TP25,26) | 0.03993118 | 0 | 0.099 | Pass |
| 9.1 | analýza | #OK | Pass | ||
| 10.2 | NZ1: Odběr proudu ze zdroje (TP19/DLO) | 5.989139E-05 | -0.001 | 0.0002 | Pass |
| 10.3 | NZ2: Odběr proudu ze zdroje (TP42/TP25,26) | 0.0001033969 | -0.001 | 0.0002 | Pass |
| 10.4 | NZ1: Odběr proudu ze zdroje (TP19/DLO) | 0.02592235 | 0.015 | 0.03 | Pass |
| 11.1 | DMM: Kontrola hodnoty proudu (TP25,26/NZ1-) | 0.145120894 | 0.13 | 1 | Pass |
| 11.2 | NZ1: Odběr proudu ze zdroje (TP19/DLO) | 0.02562846 | 0.015 | 0.03 | Pass |
| 12.1 | NZ2: Odběr proudu ze zdroje (TP22/TP25,26) | 0.04251172 | 0.02 | 0.05 | Pass |
| 12.2 | NZ1: Odběr proudu ze zdroje (TP19/DLO) | 2.723677E-05 | -0.001 | 0.0001 | Pass |
| 12.3 | NZ1: Odběr proudu ze zdroje (TP19/DLO) | 0.02549784 | 0.015 | 0.03 | Pass |
| 13.1 | NZ1: Odběr proudu ze zdroje (TP19/DLO) | -5.417849E-06 | -0.001 | 0.0001 | Pass |
| 14.1 | NZ1: Odběr proudu ze zdroje (TP22/TP25,26) | 0.04525388 | 0.03 | 0.06 | Pass |
| 15.1 | Test oscilátorů | 0 (0x00) | 0 (0x00) | 0 (0x00) | Pass |
| 16.1 | Test komunikace mezi U2 a U3 přes RS485 | 0 (0x00) | 0 (0x00) | 0 (0x00) | Pass |
| 17.1 | Test eeprom a ověření velikosti | 0 (0x00) | 0 (0x00) | 0 (0x00) | Pass |
| 17.2 | Test SPI FLASH | 0 (0x00) | 0 (0x00) | 255 (0xff) | Pass |
| 18.1 | Analysis: hodnota na max | True | d_VCC_ADC[3405]<=d_max[3447.62123697696] | Pass | |
| 19.1 | Analysis: hodnota na max | False | d_U5V_ADC[4095]<=d_max[3130.46918341398] | Fail |
Measurements over time
Hover for exact values · drag to zoom · double-click to reset
Readings (2)
| Date | Value | Min | Max |
|---|---|---|---|
| 2026-01-19 08:42 | 1487991 | — | — |
| 2026-02-02 06:42 | 1487991 | — | — |