Serial lookup
Search a serial number to see its full test history.
Serial 001791508
Tests9
Approved1
Failures8
Tests over time
Test history
| Date | Measurement ID | ID record | Article | Result | |
|---|---|---|---|---|---|
| 2024-09-09 08:04 | 260 | 638614586146778992-9802 | S0001055 | Fail | View |
| 2024-09-09 08:24 | 261 | 638614598350729917-2167 | S0001055 | Fail | View |
| 2024-09-23 09:16 | 509 | 638626725077389919-7829 | S0001055 | Fail | View |
| 2024-09-24 15:11 | 546 | 638627802151673719-9948 | S0001055 | Fail | View |
| 2024-12-05 15:17 | 5401 | 638690049814161171-9057 | S0001055 | Fail | View |
| 2024-12-05 15:19 | 5402 | 638690051319283715-3375 | S0001055 | Fail | View |
| 2025-09-05 11:26 | 25762 | 638926611421264013-1623 | S0001055 | Fail | View |
| 2025-11-06 17:32 | 40586 | 638980434964616394-8442 | S0001055 | Fail | View |
| 2026-07-09 15:00 | 77753 | 639192059897788338-4070 | S0001055 | Pass | View |
Test 25762
- Result
- Fail
- Date
- 2025-09-05 11:26
- Article
- S0001055
- ID record
- 638926611421264013-1623
- ID production
- 49267
- Temperature
- 27.59
- Humidity
- 46.58
- Test duration
- 54
- Error ID
- 372
- Error text
- TestSW: [ADC_PA1] sense B
Measurement lines
| Step | Name | Value | Min | Max | Result |
|---|---|---|---|---|---|
| 0.1 | Uložení čísla výrobního příkazu | 49267 | Pass | ||
| 0.2 | Uložení ID obsluhy | 1332 | Pass | ||
| 0.3 | Uložení teploty okolí | 27.59 | Pass | ||
| 0.4 | Uložení vlhkosti okolí | 46.58 | Pass | ||
| 0.5 | Uložení Id testeru (uloženo na lokálním disku) | Vez_Hokami_01-v01.01 | Pass | ||
| 0.6 | Uložení id Artiklu (produktové číslo) | S0001055 | Pass | ||
| 1.1 | DMM: Kontrola hodnoty Odporu (TP6/TP4) | 0.700459276 | < | 10 | Pass |
| 1.2 | DMM: Kontrola hodnoty Odporu (TP6/TP5) | 997.27805 | 800 | 1200 | Pass |
| 3.1 | NZ3.1: Měření napětí: Zkrat? | 28.00651 | > | 27.9 | Pass |
| 3.2 | NZ3.1: Měření proudu: Kontrola odběru | 0.0009207412 | < | 0.005 | Pass |
| 3.3 | R11.12: Zapnutí (TP42/TP43) | Pass | |||
| 3.4 | NZ3.1: Měření napětí: Zkrat? | 28.00651 | > | 27.9 | Pass |
| 3.5 | NZ3.1: Měření proudu: Kontrola odběru | 0.02420378 | < | 0.04 | Pass |
| 4.1 | DMM: Měření napětí: +BATT (TP49/TP44) | 27.9664502 | 27.9 | 28.1 | Pass |
| 4.2 | DMM: Měření napětí: +BATT (TP49/TP34) | 27.9971808 | 27.9 | 28.1 | Pass |
| 4.3 | DMM: Měření napětí: +12V (TP49/TP46) | 11.8038412 | 11 | 12.2 | Pass |
| 4.4 | DMM: Měření napětí: +5V (TP49/TP7) | 4.9912037 | 4.9 | 5.1 | Pass |
| 4.5 | DMM: Měření napětí: +3V3 (TP49/TP16) | 3.30353568 | 3.2 | 3.4 | Pass |
| 4.6 | DMM: Měření napětí: VDDA +3V3 (TP49/TP2) | 3.30352957 | 3.2 | 3.4 | Pass |
| 4.7 | DMM: Měření napětí: (TP49/TP8) | 4.67225173 | 4.57 | 4.77 | Pass |
| 4.8 | DMM: Měření napětí: (TP49/TP9) | 4.66944735 | 4.57 | 4.77 | Pass |
| 4.9 | DMM: Měření napětí: (TP49/TP10) | 4.66887403 | 4.57 | 4.77 | Pass |
| 5.1 | DMM: Měření napětí: (TP49/TP46) | 0.514065491 | < | 1 | Pass |
| 5.2 | DMM: Měření napětí: (TP49/TP46) | 11.7152705 | > | 11.5 | Pass |
| 6.1 | R9.12: Vypnutí (TP49/TP28) | Pass | |||
| 7.1 | NZ3: Měření napětí: Zkrat? | 28.00651 | > | 27.9 | Pass |
| 7.2 | NZ3: Měření proudu: Kontrola odběru | 0.02529706 | < | 0.06 | Pass |
| 7.3 | R0.6: Zapnutí Ext. Programátoru | Pass | |||
| 8.1 | TestSw: Ověření komunikace | [TYP:AA NAME:eDrive NAME2:TestSW HW:Kolecko v4 SW: | Pass | ||
| 9.1 | TestSW: [CHG_SIG (PC14)] = log1? | 16384 | 16384 | 16384 | Pass |
| 9.2 | EL1.1: Ověření proudu na zátěži -> Current? | 2.060242 | 1.6 | 2.8 | Pass |
| 9.3 | DMM: Měření napětí: Vstup U14 (TP39/TP32) | 29.6031692 | 29 | 30.1 | Pass |
| 9.4 | DMM: Měření napětí: Vstup U14 (TP39/TP33) | 11.2489573 | 10.8 | 11.5 | Pass |
| 9.5 | DMM: Měření napětí: Vstup U14 (TP39/TP40) | 1.56755377 | 1.5 | 1.8 | Pass |
| 10.1 | NZ3.1: Měření napětí: Zkrat? | 28.00651 | > | 27.9 | Pass |
| 10.2 | NZ3.1: Měření proudu: Kontrola odběru | 0.02519392 | < | 0.06 | Pass |
| 10.3 | TestSW: [ADC_PA03] | 3844 | 3667 | 3937 | Pass |
| 12.1 | TestSW: [PA15] - napájení sense ABC | Pass | |||
| 12.2 | TestSW: [HALL_A (PB4)] = log0? | 0 | 0 | 0 | Pass |
| 12.3 | TestSW: [HALL_A (PB4)] = log1? | 16 | 16 | 16 | Pass |
| 12.4 | TestSW: [HALL_B (PB5)] = log0? | 0 | 0 | 0 | Pass |
| 12.5 | TestSW: [HALL_B (PB5)] = log1? | 32 | 32 | 32 | Pass |
| 12.6 | TestSW: [HALL_C (PB0)] = log0? | 0 | 0 | 0 | Pass |
| 12.7 | TestSW: [HALL_C (PB0)] = log1? | 1 | 1 | 1 | Pass |
| 13.1 | TestSW: [CON_ID (PC13)] = log1? | 8192 | 8192 | 8192 | Pass |
| 13.10 | TestSW: [BTN_FFWD][ADC_PA6]? | 4086 | 4000 | 4096 | Pass |
| 13.11 | TestSW: [BTN_FFWD][ADC_PA6]? | 2026 | 1900 | 2100 | Pass |
| 13.12 | TestSW: [BTN_STOP][ADC_PA5]? | 4086 | 4000 | 4096 | Pass |
| 13.13 | TestSW: [BTN_STOP][ADC_PA5]? | 2025 | 1900 | 2100 | Pass |
| 13.2 | TestSW: [CON_ID (PC13)] = log0? | 0 | 0 | 0 | Pass |
| 13.3 | TestSW: [BTN_DRV][PA11] na log1 | Pass | |||
| 13.4 | TestSW: [BTN_FWD][ADC_PB1]? | 4085 | 4000 | 4096 | Pass |
| 13.5 | TestSW: [BTN_FWD][ADC_PB1]? | 2014 | 1900 | 2100 | Pass |
| 13.6 | TestSW: [BTN_REV][ADC_PA4]? | 4085 | 4000 | 4096 | Pass |
| 13.7 | TestSW: [BTN_REV][ADC_PA4]? | 2024 | 1900 | 2100 | Pass |
| 13.8 | TestSW: [BTN_ULCK][ADC_PA7]? | 4085 | 4000 | 4096 | Pass |
| 13.9 | TestSW: [BTN_ULCK][ADC_PA7]? | 2023 | 1900 | 2100 | Pass |
| 14.1 | TestSW: [LED1][PB2] na log1 | Pass | |||
| 14.10 | DMM: Měření napětí: LED3 (TP49/TP14) | 3.2988681 | > | 3 | Pass |
| 14.11 | TestSW: [LED3][PB11] na log0 | Pass | |||
| 14.12 | DMM: Měření napětí: LED3 (TP49/TP14) | 0.00348682558 | < | 0.1 | Pass |
| 14.13 | TestSW: [LED4][PB12] na log1 | Pass | |||
| 14.14 | DMM: Měření napětí: LED4 (TP49/TP12) | 3.30012359 | > | 3 | Pass |
| 14.15 | TestSW: [LED4][PB12] na log0 | Pass | |||
| 14.16 | DMM: Měření napětí: LED4 (TP49/TP12) | 0.00347566909 | < | 0.1 | Pass |
| 14.2 | DMM: Měření napětí: LED1 (TP49/TP13) | 3.29874994 | > | 3 | Pass |
| 14.3 | TestSW: [LED1][PB2] na log0 | Pass | |||
| 14.4 | DMM: Měření napětí: LED1 (TP49/TP13) | 0.00367740278 | < | 0.1 | Pass |
| 14.5 | TestSW: [LED2][PB10] na log1 | Pass | |||
| 14.6 | DMM: Měření napětí: LED1 (TP49/TP11) | 3.29885637 | > | 3 | Pass |
| 14.7 | TestSW: [LED2][PB10] na log0 | Pass | |||
| 14.8 | DMM: Měření napětí: LED1 (TP49/TP11) | 0.00353530831 | < | 0.1 | Pass |
| 14.9 | TestSW: [LED3][PB11] na log1 | Pass | |||
| 15.1 | TestSW: ntc | 54 | 49 | 90 | Pass |
| 16.1 | DMM: Měření frekvence (TP49/TP55) | 19995.1098 | 19600 | 20400 | Pass |
| 16.2 | DMM: Měření frekvence (TP49/TP56) | 19995.108 | 19600 | 20400 | Pass |
| 16.3 | DMM: Měření napětí (TP49/TP52) | 14.0121528 | 13.3 | 14.7 | Pass |
| 17.1 | DMM: Měření frekvence (TP49/TP61) | 19995.1027 | 19600 | 20400 | Pass |
| 17.2 | DMM: Měření frekvence (TP49/TP62) | 19995.1065 | 19600 | 20400 | Pass |
| 17.3 | DMM: Měření napětí (TP49/TP58) | 14.0133783 | 13.3 | 14.7 | Pass |
| 18.1 | DMM: Měření frekvence (TP49/TP67) | 19995.1009 | 19600 | 20400 | Pass |
| 18.2 | DMM: Měření frekvence (TP49/TP68) | 19995.1042 | 19600 | 20400 | Pass |
| 18.3 | DMM: Měření napětí (TP49/TP58) | 14.0120907 | 13.3 | 14.7 | Pass |
| 19.1 | DMM: Měření napětí: +5V (TP49/TP7) | 0.00230610842 | < | 0.5 | Pass |
| 20.0.3 | TestSW: [ADC_PA2] sense C do var | VC = 2261 | Pass | ||
| 20.0.7 | Analysis: VC | VC[2261] | 2210 | 2340 | Pass |
| 20.0.6 | Analysis: VB | VB[2256] | 2210 | 2340 | Pass |
| 20.0.5 | Analysis: VA | VA[2265] | 2210 | 2340 | Pass |
| 20.0.2 | TestSW: [ADC_PA1] sense B do var | VB = 2256 | Pass | ||
| 20.0.1 | TestSW: [ADC_PA0] sense A do var | VA = 2265 | Pass | ||
| 20.1 | NZ3.1: Měření napětí: Zkrat? | 28.00651 | > | 27.9 | Pass |
| 20.1.1 | EL1.2: Ověření proudu na zátěži -> Current? | 4.989916 | 4.7 | 5.3 | Pass |
| 20.1.2 | TestSW: [ADC_PA0] sense A | 2436 | 2430[VA] | 2530[VA] | Pass |
| 20.1.3 | TestSW: [ADC_PA1] sense B | 2091 | 1991[VB] | 2091[VB] | Pass |
| 20.1.4 | TestSW: [ADC_PA2] sense C | 2262 | 2210 | 2340 | Pass |
| 20.2 | NZ3.1: Měření proudu: Kontrola odběru | 0.03773276 | < | 0.06 | Pass |
| 20.2.1 | EL1.2: Ověření proudu na zátěži -> Current? | 4.989839 | 4.7 | 5.3 | Pass |
| 20.2.2 | TestSW: [ADC_PA0] sense A | 2093 | 2000[VA] | 2100[VA] | Pass |
| 20.2.3 | TestSW: [ADC_PA1] sense B | 2420 | 2421[VB] | 2521[VB] | Fail |
Measurements over time
Hover for exact values · drag to zoom · double-click to reset
Readings (1)
| Date | Value | Min | Max |
|---|---|---|---|
| 2026-07-09 15:00 | 1.047 | — | 10 |