Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00R00004

Tests4
Approved1
Failures3

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-12 14:19 23699 639011459818336512-5655 S0003591 Fail View
2025-12-12 14:20 23700 639011460299561506-3119 S0003591 Fail View
2025-12-12 14:21 23701 639011460860053819-6463 S0003591 Fail View
2025-12-12 14:24 23703 639011462384094656-4235 S0003591 Pass View

Test 23701

Result
Fail
Date
2025-12-12 14:21
Article
S0003591
ID record
639011460860053819-6463
ID production
Temperature
Humidity
Test duration
28
Error ID
218
Error text
TestSW: test oscilátoru ON

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5595.76086 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 999.03134 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.681314208 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.892545478 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1427622.2 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.15300795 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.28518274 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.27751478 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.27580469 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.80192399 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9914158 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0162461 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9813135 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (4)
DateValueMinMax
2025-12-12 14:19 999.112 950 1050
2025-12-12 14:20 999.074 950 1050
2025-12-12 14:21 999.031 950 1050
2025-12-12 14:24 999.056 950 1050