Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000155

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-12 13:40 23664 639011436053420235-6350 S0003591 Fail View
2025-12-12 13:41 23665 639011436632316414-3865 S0003591 Fail View
2025-12-12 13:42 23666 639011437107448857-5686 S0003591 Pass View

Test 23665

Result
Fail
Date
2025-12-12 13:41
Article
S0003591
ID record
639011436632316414-3865
ID production
Temperature
Humidity
Test duration
17
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5595.64693 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1000.21278 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.641545948 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.625417534 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1363550.78 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.20701134 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.28700483 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.2797884 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.27781691 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79947092 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.992506 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0090588 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9926121 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2025-12-12 13:40 1000.276 950 1050
2025-12-12 13:41 1000.213 950 1050
2025-12-12 13:42 1000.176 950 1050