Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000201

Tests4
Approved1
Failures3

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-12 12:29 23603 639011393167468008-1441 S0003591 Fail View
2025-12-12 12:30 23604 639011393701448088-4029 S0003591 Fail View
2025-12-12 12:34 23608 639011396595632330-7442 S0003591 Fail View
2025-12-12 12:35 23609 639011397011725222-3796 S0003591 Pass View

Test 23608

Result
Fail
Date
2025-12-12 12:34
Article
S0003591
ID record
639011396595632330-7442
ID production
Temperature
Humidity
Test duration
16
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5615.64884 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1001.42826 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.640175494 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.39333224 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1424043.86 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.1909681 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29341664 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28598339 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.28417259 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79752941 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9889619 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00794234 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9517924 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (4)
DateValueMinMax
2025-12-12 12:29 1001.528 950 1050
2025-12-12 12:30 1001.438 950 1050
2025-12-12 12:34 1001.428 950 1050
2025-12-12 12:35 1001.363 950 1050