Serial C00000209

Tests4
Approved1
Failures3

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-12 12:14 23590 639011384526263243-3469 S0003591 Fail View
2025-12-12 12:15 23591 639011385087809429-6312 S0003591 Fail View
2025-12-12 12:16 23592 639011385575731277-1176 S0003591 Fail View
2025-12-12 12:17 23593 639011386105361609-4294 S0003591 Pass View

Test 23590

Result
Fail
Date
2025-12-12 12:14
Article
S0003591
ID record
639011384526263243-3469
ID production
Temperature
Humidity
Test duration
27
Error ID
218
Error text
TestSW: test oscilátoru ON

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5622.11629 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 999.053089 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.432787936 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.635101201 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1212604.34 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.15392442 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29514233 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28868801 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.29373976 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.80039681 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.984548 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0181187 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9642418 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (4)
DateValueMinMax
2025-12-12 12:14 999.053 950 1050
2025-12-12 12:15 998.991 950 1050
2025-12-12 12:16 998.89 950 1050
2025-12-12 12:17 998.802 950 1050