Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000114

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-12 11:00 23525 639011339621416355-3136 S0003591 Fail View
2025-12-12 11:00 23526 639011340111527526-4929 S0003591 Fail View
2025-12-12 11:01 23527 639011340525771813-4041 S0003591 Pass View

Test 23526

Result
Fail
Date
2025-12-12 11:00
Article
S0003591
ID record
639011340111527526-4929
ID production
Temperature
Humidity
Test duration
17
Error ID
208
Error text
Vyhodnocení programu: [testSW]

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5613.30636 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1000.83892 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.647211541 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.618441998 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1452230.41 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.20311171 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29011198 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28232836 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.28091281 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79878823 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9773443 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00767916 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.89935 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2025-12-12 11:00 1000.94 950 1050
2025-12-12 11:00 1000.839 950 1050
2025-12-12 11:01 1000.805 950 1050