Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000261

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-12 09:06 23432 639011271292815031-7547 S0003591 Fail View
2025-12-12 09:06 23433 639011271773615784-9025 S0003591 Fail View
2025-12-12 09:07 23434 639011272251666574-3290 S0003591 Pass View

Test 23433

Result
Fail
Date
2025-12-12 09:06
Article
S0003591
ID record
639011271773615784-9025
ID production
Temperature
Humidity
Test duration
23
Error ID
219
Error text
DMM: Měření frekvence: 1kHz (TP2/TP17)

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5601.08872 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 998.473571 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.638040018 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.830977576 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1382839.27 120000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.19798191 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.28324959 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.27526945 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.27350714 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79583167 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9895788 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00777069 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9567173 11.4 12.6 Pass
3.2.1 DMM: Měření frekvence: 1kHz (TP2/TP17) 109.829676 999.8 1000.2 Fail

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2025-12-12 09:06 998.595 950 1050
2025-12-12 09:06 998.474 950 1050
2025-12-12 09:07 998.438 950 1050