Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial CR00000061

Tests6
Approved1
Failures5

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-12-04 10:45 22993 639004419356662851-9246 S0003591 Fail View
2025-12-04 10:46 22994 639004419803688772-9055 S0003591 Fail View
2025-12-04 10:51 22995 639004422510265943-4968 S0003591 Fail View
2025-12-04 11:03 22996 639004430080089078-1405 S0003591 Fail View
2025-12-04 11:24 22997 639004442400628494-9503 S0003591 Fail View
2025-12-04 11:26 22998 639004442964141629-5290 S0003591 Pass View

Test 22994

Result
Fail
Date
2025-12-04 10:46
Article
S0003591
ID record
639004419803688772-9055
ID production
Temperature
Humidity
Test duration
6
Error ID
120
Error text
DMM: Měření odporu: R3+R4 (TP10/TP8)

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 4696.98319 4465 4935 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 2199.61372 2090 2310 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.650498376 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.859629234 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 870875.158 1200000 1540000 Fail

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (6)
DateValueMinMax
2025-12-04 10:45 2199.62 2090 2310
2025-12-04 10:46 2199.614 2090 2310
2025-12-04 10:51 2199.618 2090 2310
2025-12-04 11:03 2199.612 2090 2310
2025-12-04 11:24 3293.067 700 6800
2025-12-04 11:26 3293.014 700 6800