Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000060

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-11-18 09:46 22875 638990559734416559-2777 S0003591 Fail View
2025-11-18 09:46 22876 638990559890293759-9079 S0003591 Fail View
2025-11-18 09:48 22877 638990560482790638-6378 S0003591 Pass View

Test 22876

Result
Fail
Date
2025-11-18 09:46
Article
S0003591
ID record
638990559890293759-9079
ID production
Temperature
Humidity
Test duration
23
Error ID
214
Error text
TestSW: test oscilátoru ON

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5598.81624 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1002.43461 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.575300877 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.59786277 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1344922.13 1200000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.14876438 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29224408 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28586023 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.29083681 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79858573 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9952291 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0108054 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9743668 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2025-11-18 09:46 1002.428 950 1050
2025-11-18 09:46 1002.435 950 1050
2025-11-18 09:48 1002.418 950 1050