Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000018

Tests3
Approved2
Failures1

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-11-18 09:42 22872 638990557216231746-3108 S0003591 Pass View
2025-11-18 09:44 22873 638990558179300995-1550 S0003591 Fail View
2025-11-18 09:45 22874 638990558741032384-7622 S0003591 Pass View

Test 22873

Result
Fail
Date
2025-11-18 09:44
Article
S0003591
ID record
638990558179300995-1550
ID production
Temperature
Humidity
Test duration
23
Error ID
214
Error text
TestSW: test oscilátoru ON

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5620.32262 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1000.56362 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.611897049 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.803768747 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1242999.26 1200000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.15043065 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29265962 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28705749 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.29118882 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79735235 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9857627 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.0086893 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9599959 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2025-11-18 09:42 999.834 950 1050
2025-11-18 09:44 1000.564 950 1050
2025-11-18 09:45 1000.531 950 1050