Serial lookup
Search a serial number to see its full test history.
Serial C00000020
Tests3
Approved1
Failures2
Tests over time
Test history
| Date | Measurement ID | ID record | Article | Result | |
|---|---|---|---|---|---|
| 2025-11-18 09:31 | 22865 | 638990550750572123-4202 | S0003591 | Fail | View |
| 2025-11-18 09:31 | 22866 | 638990550903197005-4076 | S0003591 | Fail | View |
| 2025-11-18 09:33 | 22867 | 638990551430887875-5251 | S0003591 | Pass | View |
Test 22865
- Result
- Fail
- Date
- 2025-11-18 09:31
- Article
- S0003591
- ID record
- 638990550750572123-4202
- ID production
- Temperature
- Humidity
- Test duration
- 5
- Error ID
- 120
- Error text
- DMM: Měření odporu: R3+R4 (TP10/TP8)
Measurement lines
| Step | Name | Value | Min | Max | Result |
|---|---|---|---|---|---|
| 1.1.1 | DMM: Měření odporu: R1 (TP5/TP4) | 5602.61 | 5320 | 5880 | Pass |
| 1.1.2 | DMM: Měření odporu: R2 (TP5/TP6) | 998.574388 | 950 | 1050 | Pass |
| 1.2.1 | DMM: Měření odporu: F1 (TP9/TP8) | 0.607266901 | < | 2 | Pass |
| 1.2.2 | DMM: Měření odporu: F2 (TP10/TP11) | 0.793227687 | < | 2.5 | Pass |
| 1.2.3 | DMM: Měření odporu: R3+R4 (TP10/TP8) | 688127.72 | 1200000 | 1540000 | Fail |
Measurements over time
Hover for exact values · drag to zoom · double-click to reset
Readings (3)
| Date | Value | Min | Max |
|---|---|---|---|
| 2025-11-18 09:31 | 998.574 | 950 | 1050 |
| 2025-11-18 09:31 | 998.566 | 950 | 1050 |
| 2025-11-18 09:33 | 998.566 | 950 | 1050 |