Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000028

Tests3
Approved1
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-11-18 09:09 22847 638990537368506378-8868 S0003591 Fail View
2025-11-18 09:09 22848 638990537537543401-5671 S0003591 Fail View
2025-11-18 09:11 22849 638990538222187661-9418 S0003591 Pass View

Test 22848

Result
Fail
Date
2025-11-18 09:09
Article
S0003591
ID record
638990537537543401-5671
ID production
Temperature
Humidity
Test duration
23
Error ID
214
Error text
TestSW: test oscilátoru ON

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5596.01608 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1001.10127 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.811039371 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 1.01781648 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 1404450.38 1200000 1540000 Pass
2.2.1 DMM: Měření napětí: +5V StepDown Output (TP2/TP13) 5.1391982 4 6 Pass
2.2.2 DMM: Měření napětí: +3V3 LDO Output (TP2/TP14) 3.29043034 2.64 3.96 Pass
2.2.3 DMM: Měření napětí: +3V3 Modem AVCC (TP2/TP16) 3.28497588 2.64 3.96 Pass
2.2.4 DMM: Měření napětí: +3V3 VDDA (TP2/TP22) 3.28908946 2.64 3.96 Pass
2.2.5 DMM: Měření napětí: +1V8 Audio (TP2/TP24) 1.79619234 1.44 2.16 Pass
2.2.6 DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28) 11.9947098 9.6 14.4 Pass
2.3.1 AC1: Měření proudu: Kontrola odběru 0.00858201 < 0.12 Pass
2.3.2 DMM: Měření napětí: +12V ACDC měnič (TP2/TP12) 11.9193272 11.4 12.6 Pass

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (3)
DateValueMinMax
2025-11-18 09:09 1001.097 950 1050
2025-11-18 09:09 1001.101 950 1050
2025-11-18 09:11 1001.086 950 1050