Testing Portal everyone (auth disabled)

Serial lookup

Search a serial number to see its full test history.

Serial C00000048

Tests4
Approved2
Failures2

Tests over time

Test history

DateMeasurement IDID recordArticleResult
2025-11-18 08:34 22821 638990516834440928-6395 S0003591 Fail View
2025-11-18 08:36 22822 638990517570282594-9712 S0003591 Pass View
2025-11-18 08:37 22823 638990518390644746-1352 S0003591 Fail View
2025-11-18 08:40 22824 638990519615240066-9503 S0003591 Pass View

Test 22823

Result
Fail
Date
2025-11-18 08:37
Article
S0003591
ID record
638990518390644746-1352
ID production
Temperature
Humidity
Test duration
5
Error ID
120
Error text
DMM: Měření odporu: R3+R4 (TP10/TP8)

Measurement lines

StepNameValueMinMaxResult
1.1.1 DMM: Měření odporu: R1 (TP5/TP4) 5618.36536 5320 5880 Pass
1.1.2 DMM: Měření odporu: R2 (TP5/TP6) 1000.5617 950 1050 Pass
1.2.1 DMM: Měření odporu: F1 (TP9/TP8) 0.597496887 < 2 Pass
1.2.2 DMM: Měření odporu: F2 (TP10/TP11) 0.558143248 < 2.5 Pass
1.2.3 DMM: Měření odporu: R3+R4 (TP10/TP8) 689346.736 1200000 1540000 Fail

Measurements over time

Hover for exact values · drag to zoom · double-click to reset

Readings (4)
DateValueMinMax
2025-11-18 08:34 1000.048 950 1050
2025-11-18 08:36 1000.04 950 1050
2025-11-18 08:37 1000.562 950 1050
2025-11-18 08:40 1000.555 950 1050