Serial lookup
Search a serial number to see its full test history.
Serial 001739099
Tests2
Approved1
Failures1
Tests over time
Test history
| Date | Measurement ID | ID record | Article | Result | |
|---|---|---|---|---|---|
| 2025-06-20 08:53 | 19693 | 638859991569173716-2255 | S0003074 | Fail | View |
| 2026-02-17 09:21 | 63450 | 639069132247241663-2398 | S0003074 | Pass | View |
Test 19693
- Result
- Fail
- Date
- 2025-06-20 08:53
- Article
- S0003074
- ID record
- 638859991569173716-2255
- ID production
- 52071
- Temperature
- 26.9
- Humidity
- 33.73
- Test duration
- 62
- Error ID
- 153
- Error text
- DMM: Měření napětí: (TP6/TP4)
Measurement lines
| Step | Name | Value | Min | Max | Result |
|---|---|---|---|---|---|
| 0.3 | Uložení teploty okolí | 26.9 | Pass | ||
| 0.4 | Uložení vlhkosti okolí | 33.73 | Pass | ||
| 0.5 | Uložení Id testeru (uloženo na lokálním disku) | Vez_Hokami_01-v01.01 | Pass | ||
| 0.6 | Uložení id Artiklu (produktové číslo) | S0003074 | Pass | ||
| 1.1 | DMM: Měření napětí: (TP2/TP4) | 0.121401372 | 5.9 | 6.3 | Fail |
| 1.1 | DMM: Měření napětí: (TP2/TP4) | 6.07846232 | 5.9 | 6.3 | Pass |
| 1.1 | DMM: Měření proudu: (TP100/TP101) změřít proud naprázdno | 0.000132048377 | 8E-05 | 0.00017 | Pass |
| 1.1 | DMM: Měření napětí: (TP2/TP4) | 5.96902992 | 5.8 | 6.3 | Pass |
| 1.1 | DMM: Měření napětí: (TP8/TP4) | 3.36643469 | 2.6 | 3.6 | Pass |
| 1.1 | DMM: Měření napětí: (TP6/TP4) | 9.01088854 | 8.6 | 9 | Fail |
| 9.2 | EL1.1: TP2/TP5 -> Current? | 0.4998813 | 0.475 | 0.525 | Pass |
Measurements over time
Hover for exact values · drag to zoom · double-click to reset
Readings (1)
| Date | Value | Min | Max |
|---|---|---|---|
| 2026-02-17 09:21 | 1739099 | — | — |