Testing Portal everyone (auth disabled)

← Portfolio

SB-VBB-050_P04

Diagnostics

Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.

Critical

Pass rate far below target

Pass rate is 50.89% (229 of 450 tests passed, 221 failed) against a target of 98%.

Warning

One measurement step drives the failures

Measurement step 12.1 — "Vyhodnocení programu: [11.0_FSW]" failed 215 times across the window's 221 failed tests.

Warning

5 serial number(s) failed repeatedly

Worst unit 001594657 failed 14 times (last on 2025-09-23 12:40). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 001594657 →

Info

No historical comparison

Historical comparison is not available for the selected window (no tests in the equally long window before it).

Info

Most frequent failure mode

Error 108 — "DMM: Měření napětí: I2C_SCL (TP41/TP8)" occurred 32 times, 14.5% of all 221 failures in the window.

Repeatedly failing units

SerialFailed testsLast failure
001594657 14 2025-09-23 12:40 Open
001417063 8 2025-01-31 09:58 Open
001417060 7 2025-11-12 11:13 Open
001416992 7 2025-09-17 14:23 Open
001594700 7 2025-09-17 14:14 Open