S0004448

Diagnostics

Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.

Critical

Pass rate far below target

Pass rate is 0% (0 of 4 tests passed, 4 failed) against a target of 98%.

Warning

One failure mode dominates

Error 174 — "DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28)" occurred 2 times, 50% of all 4 failures in the window.

Warning

One measurement step drives the failures

Measurement step 2.2.6 — "DMM: Měření napětí: +12V Bed Interface Power (TP2/TP28)" failed 2 times across the window's 4 failed tests.

Warning

1 serial number(s) failed repeatedly

Worst unit C02000033 failed 4 times (last on 2026-06-19 10:32). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect C02000033 →

Info

No historical comparison

Historical comparison is not available for the selected window (no tests in the equally long window before it).

Repeatedly failing units

SerialFailed testsLast failure
C02000033 4 2026-06-19 10:32 Open