Testing Portal everyone (auth disabled)

← Portfolio

S0004028

Diagnostics

Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.

Critical

Pass rate far below target

Pass rate is 75.38% (949 of 1,259 tests passed, 310 failed) against a target of 98%.

Warning

One failure mode dominates

Error 191 — "NZ1: Odběr proudu ze zdroje (TP19/TP25,26)" occurred 187 times, 60.3% of all 310 failures in the window.

Warning

One measurement step drives the failures

Measurement step 7.1 — "NZ1: Odběr proudu ze zdroje (TP19/TP25,26)" failed 1,231 times across the window's 310 failed tests.

Warning

5 serial number(s) failed repeatedly

Worst unit 002288409 failed 6 times (last on 2025-11-19 10:08). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 002288409 →

Info

No historical comparison

Historical comparison is not available for the selected window (no tests in the equally long window before it).

Repeatedly failing units

SerialFailed testsLast failure
002288409 6 2025-11-19 10:08 Open
001607571 4 2025-11-20 20:37 Open
001607445 4 2025-11-20 10:01 Open
001573981 3 2025-11-21 11:28 Open
001573931 3 2025-11-21 11:28 Open