Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.

Critical

Pass rate far below target

Pass rate is 82.44% (108 of 131 tests passed, 23 failed) against a target of 98%.

Warning

One failure mode dominates

Error 060 — "DMM: Kontrola hodnoty Odporu (TP4/TP3)" occurred 14 times, 60.9% of all 23 failures in the window.

Warning

One measurement step drives the failures

Measurement step 8.1 — "Vyhodnocení programu: [testSW]" failed 111 times across the window's 23 failed tests.

Warning

4 serial number(s) failed repeatedly

Worst unit 001359908 failed 5 times (last on 2025-12-09 13:42). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 001359908 →

Info

No historical comparison

Historical comparison is not available for the selected window (no tests in the equally long window before it).

Repeatedly failing units

SerialFailed testsLast failure
001359908 5 2025-12-09 13:42 Open
001359802 4 2025-12-09 13:18 Open
001359799 3 2025-12-09 10:01 Open
001359922 2 2025-12-09 10:23 Open