S0004022
Diagnostics
Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.
Pass rate far below target
Pass rate is 5.88% (3 of 51 tests passed, 48 failed) against a target of 98%.
One measurement step drives the failures
Measurement step 1.0.4 — "DMM: Měření proudu: sloupec A (TP16/TP3)" failed 18 times across the window's 48 failed tests.
5 serial number(s) failed repeatedly
Worst unit 002342288 failed 11 times (last on 2026-07-14 15:00). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 002342288 →
No historical comparison
Historical comparison is not available for the selected window (no tests in the equally long window before it).
Most frequent failure mode
Error 106 — "DMM: Měření proudu: sloupec A (TP16/TP3)" occurred 18 times, 37.5% of all 48 failures in the window.