S0003590
Diagnostics
Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.
Pass rate far below target
Pass rate is 66.74% (1,465 of 2,195 tests passed, 730 failed) against a target of 98%.
One failure mode dominates
Error 270 — "TestSW: commPLM test" occurred 393 times, 53.8% of all 730 failures in the window.
5 serial number(s) failed repeatedly
Worst unit C01000352 failed 23 times (last on 2026-07-15 04:13). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect C01000352 →
No historical comparison
Historical comparison is not available for the selected window (no tests in the equally long window before it).
Most-failing measurement step
Measurement step 3.4.8 — "DMM: Měření odporu: (TP39/TP40)" failed 65 times across the window's 730 failed tests.