Testing Portal everyone (auth disabled)

← Portfolio

S0003075

Diagnostics

Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.

Critical

Pass rate far below target

Pass rate is 87.94% (372 of 423 tests passed, 51 failed) against a target of 98%.

Warning

5 serial number(s) failed repeatedly

Worst unit 001943912 failed 9 times (last on 2025-10-09 06:47). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 001943912 →

Info

No historical comparison

Historical comparison is not available for the selected window (no tests in the equally long window before it).

Info

Most frequent failure mode

Error 80 — "NZ3: Měření proudu: Kontrola odběru" occurred 25 times, 49% of all 51 failures in the window.

Info

Most-failing measurement step

Measurement step 12.3 — "EL: Měření napětí LED3: (TP1/TP31)" failed 5 times across the window's 51 failed tests.

Repeatedly failing units

SerialFailed testsLast failure
001943912 9 2025-10-09 06:47 Open
S000307500 8 2024-10-29 12:33 Open
001610104 7 2024-11-27 08:12 Open
001610102 7 2024-10-29 12:45 Open
001574507 5 2025-10-09 06:52 Open