S0001055
Diagnostics
Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.
Pass rate far below target
Pass rate is 61.45% (1,422 of 2,314 tests passed, 892 failed) against a target of 98%.
5 serial number(s) failed repeatedly
Worst unit 001791579 failed 26 times (last on 2026-09-23 05:45). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 001791579 →
No historical comparison
Historical comparison is not available for the selected window (no tests in the equally long window before it).
Most frequent failure mode
Error 142 — "DMM: Měření napětí: (TP49/TP46)" occurred 160 times, 17.9% of all 892 failures in the window.
Most-failing measurement step
Measurement step 5.2 — "DMM: Měření napětí: (TP49/TP46)" failed 167 times across the window's 892 failed tests.