S0000608B
Diagnostics
Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.
Pass rate far below target
Pass rate is 0% (0 of 14 tests passed, 14 failed) against a target of 98%.
One measurement step drives the failures
Measurement step 5.1 — "DMM: Kontrola hodnoty napětí (TP12/TP17,18)" failed 5 times across the window's 14 failed tests.
4 serial number(s) failed repeatedly
Worst unit 001875979 failed 6 times (last on 2024-10-17 15:43). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 001875979 →
No historical comparison
Historical comparison is not available for the selected window (no tests in the equally long window before it).
Most frequent failure mode
Error 118 — "DMM: Kontrola hodnoty napětí (TP12/TP17,18)" occurred 5 times, 35.7% of all 14 failures in the window.