Testing Portal everyone (auth disabled)

← Portfolio

S0000608

Diagnostics

Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.

Warning

Pass rate below target

Pass rate is 95.82% (2,752 of 2,872 tests passed, 120 failed) against a target of 98%.

Warning

One measurement step drives the failures

Measurement step 29.1 — "Vyhodnocení programu: [finalSW]" failed 2,050 times across the window's 120 failed tests.

Warning

5 serial number(s) failed repeatedly

Worst unit 001856840 failed 6 times (last on 2025-07-23 14:04). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 001856840 →

Info

No historical comparison

Historical comparison is not available for the selected window (no tests in the equally long window before it).

Info

Most frequent failure mode

Error 57 — "DMM: Kontrola hodnoty Odporu (TP2/TP3)" occurred 33 times, 27.5% of all 120 failures in the window.

Repeatedly failing units

SerialFailed testsLast failure
001856840 6 2025-07-23 14:04 Open
002200178 5 2025-08-15 09:01 Open
001129235 5 2024-11-26 13:01 Open
001342240 5 2024-09-24 13:56 Open
001342222 5 2024-09-24 13:45 Open