S0000608
Diagnostics
Deterministic checks over the selected window — every statement is computed from the data, nothing is generated.
Pass rate below target
Pass rate is 95.82% (2,752 of 2,872 tests passed, 120 failed) against a target of 98%.
One measurement step drives the failures
Measurement step 29.1 — "Vyhodnocení programu: [finalSW]" failed 2,050 times across the window's 120 failed tests.
5 serial number(s) failed repeatedly
Worst unit 001856840 failed 6 times (last on 2025-07-23 14:04). Repeatedly failing units usually indicate a real hardware defect rather than a flaky test.Inspect 001856840 →
No historical comparison
Historical comparison is not available for the selected window (no tests in the equally long window before it).
Most frequent failure mode
Error 57 — "DMM: Kontrola hodnoty Odporu (TP2/TP3)" occurred 33 times, 27.5% of all 120 failures in the window.